120
c
HAPTER
6:
Computer Interface Operation
Model 335 Temperature Controller
MODE
Remote Interface Mode Command
Input
MODE <mode>[term]
Format
n
<mode>
0 = local, 1 = remote, 2 = remote with local lockout.
Example
MODE 2[term]
places the Model 335 into remote mode with local lockout.
MODE?
Remote Interface Mode Query
Input
MODE?[term]
Returned
<mode>[term]
Format
n (refer to command for description)
MOUT
Manual Output Command
Input
MOUT <output>,<value>[term]
Format
n,
+
nnnnn[term]
<output>
Specifies output to configure: 1 or 2.
<value>
Specifies value for manual output.
Example
MOUT 1,22.45[term]
Output 1 manual output is 22.45%.
Remarks
Manual output only applies to outputs in Closed Loop PID, Zone, or Open Loop modes.
MOUT?
Manual Output Query
Input
MOUT? <output>[term]
Format
n
<output>
Specifies which output to query: 1 or 2.
Returned
<value>
Format
+
nnnnn[term] (refer to command for description)
OPST?
Operational Status Query
Input
OPST? [term]
Returned
<bit weighting> [term]
Format
nnn
Remarks
The integer returned represents the sum of the bit weighting of the operational sta-
tus bits. Refer to section 6.2.5.2 for a list of operational status bits.
OPSTE
Operational Status Enable Command
Input
OPSTE <bit weighting> [term]
Format
nnn
Remarks
Each bit has a bit weighting and represents the enable/disable mask of the corre-
sponding operational status bit in the Operational Status Register. This determines
which status bits can set the corresponding summary bit in the Status Byte Register.
To enable a status bit, send the command OPSTE with the sum of the bit weighting for
each desired bit. Refer to section 6.2.5.2 for a list of operational status bits.
OPSTE?
Operational Status Enable Query
Input
OPSTE?[term]
Returned
<bit weighting> [term]
Format
nnn (Refer to section 6.2.5.2 for a list of operational status bits)