HTRSET
Heater Setup Command
Input
HTRSET <output>,<type>,<heater resistance>,<max cur-
rent>,<max user current>,<current/power>[term]
Format
n,n,n,n,+n.nnn,n
<output>
Specifies which heater output to configure: 1 or 2.
<type>
Output type (Output 2 only): 0=Current, 1=Voltage
<htr resistance>
Heater Resistance Setting: 1 = 25
)
, 2 = 50
)
.
<max current>
Specifies the maximum heater output current:
0 = User Specified, 1 = 0.707 A, 2 = 1 A, 3 = 1.141 A,
4 = 1.732 A
<max user current>
Specifies the maximum heater output current if
max current is set to User Specified.
<current/power>
Specifies whether the heater output displays in current or
power (current mode only). Valid entries: 1 = current,
2 = power.
Example
HTRSET 2,0,1,2,0,1[term]
Heater Output 2 uses the current source output, the
25
)
heater setting, has a maximum current of 1 A, the maximum user current is set
to 0 A because it is not going to be used since a discrete value has been chosen, and
the heater output will be displayed in units of current.
Remarks
Max current will be limited to 1 A on Output 1 if the heater resistance is set to 50
)
. If
the heater resistance is set to 25
)
on Output 1, then the max current will be limited
to 1.414 A if Output 2 is in Current mode, or 1.732 A if Output 2 is in Voltage mode. See
section 4.5.1.3 for more information on the max current setting.
HTRSET?
Heater Setup Query
Input
HTRSET? <output>[term]
Format
n
<output>
Specifies which heater output to query: 1 or 2.
Returned
<type>,<htr resistance>,<max current>,<max user current>,<current/power>[term]
Format
n,n,n,+n.nnn,n
HTRST?
Heater Status Query
Input
HTRST? <output>[term]
Format
n
<output>
Specifies which heater output to query: 1 or 2.
Returned
<error code>[term]
Format
n
<error code> Heater error code: 0 = no error, 1 = heater open load, 2 = heater short.
Remarks
Error condition is cleared upon querying the heater status, which will also clear the
front panel error message.
IEEE
IEEE-488 Interface Parameter Command
Input
IEEE <address>[term]
Format
nn
<address>
Specifies the IEEE address: 1–30. (Address 0 and 31 are reserved.)
Example
IEEE 4[term]
after receipt of the current terminator, the instrument responds to
address 4.