112
c
HAPTER
6:
Computer Interface Operation
Model 335 Temperature Controller
BRIGT?
Display Brightness Query
Input
BRIGT?[term]
Returned
<brightness value>[term]
Format
n (refer to command for description)
CRDG?
Celsius Reading Query
Input
CRDG? <input>[term]
Format
a
<input> A or B
Returned
<temp value>[term]
Format
±nnnnnn
Remarks
Also see the RDGST? query.
CRVDEL
Curve Delete Command
Input
CRVDEL <curve>[term]
Format
nn
<curve> Specifies a user curve to delete. Valid entries: 21–59.
Example
CRVDEL 21[term]
deletes User Curve 21.
CRVHDR
Curve Header Command
Input
CRVHDR <curve>,<name>,<SN>,<format>,<limit value>,<coeffi-
cient>[term]
Format
nn,s[15],s[10],n,+nnn.nnn,n
<curve>
Specifies which curve to configure. Valid entries: 21–59.
<name>
Specifies curve name. Limited to 15 characters.
<SN>
Specifies the curve serial number. Limited to ten characters.
<format>
Specifies the curve data format. Valid entries: 1 = mV/K, 2 = V/K,
3 =
)
/K, 4 = log
)
/K.
<limit value> Specifies the curve temperature limit in kelvin.
<coefficient> Specifies the curves temperature coefficient. Valid entries:
1 = negative, 2 = positive.
Remarks
Configures the user curve header. The coefficient parameter will be calculated auto-
matically based on the first two curve datapoints. It is included as a parameter for
compatability with the CRVHDR? query.
Example
CRVHDR 21,DT-470,00011134,2,325.0,1[term]
configures User Curve 21 with a
name of DT-470, serial number of 00011134, data format of volts versus kelvin, upper
temperature limit of 325 K, and negative coefficient.
CRVHDR?
Curve Header Query
Input
CRVHDR? <curve>[term]
Format
nn
<curve>
Valid entries: 1–59.
Returned
<name>,<SN>,<format>,<limit value>,<coefficient>[term]
Format
s[15],s[10],n,+nnn.nnn,n (refer to command for description)