![Lake Shore 335 Скачать руководство пользователя страница 120](http://html.mh-extra.com/html/lake-shore/335/335_user-manual_670643120.webp)
108
c
HAPTER
6:
Computer Interface Operation
Model 335 Temperature Controller
ESR?
Standard Event Status Register Query
Input
ESR?[term]
Returned
<bit weighting>
Format
nnn
Remarks
The integer returned represents the sum of the bit weighting of the event flag bits in
the Standard Event Status Register. Refer to section 6.2.5 for a list of event flags.
IDN?
Identification Query
Input
IDN?[term]
Returned
<manufacturer>,<model>,<instrument serial>/<option serial>,
<firmware version>[term]
Format
s[4],s[8],s[7]/s[7],n.n
<manufacturer>
Manufacturer ID
<model>
Instrument model number
<instrument serial>
Instrument serial number
<option card serial>
Option card serial number
<firmware version>
Instrument firmware version
Example
LSCI,MODEL335,1234567/1234567,1.0
OPC
Operation Complete Command
Input
OPC[term]
Remarks
Generates an Operation Complete event in the Event Status Register upon comple-
tion of all pending selected device operations. Send it as the last command in a com-
mand string.
OPC?
Operation Complete Query
OPC?[term]
Returned
1[term]
Remarks
Places a 1 in the controller output queue upon completion of all pending selected
device operations. Send as the last command in a command string.
Not the same as
OPC.
RST
Reset Instrument Command
Input
RST[term]
Remarks
Sets controller parameters to power-up settings.