CHAPTER 3:
Installation and Setup
Model CRX-6.5K Probe Station
3.8 System
Verification
and Testing
This section describes a sequence of short tests that can be used to verify that the
probe station has been assembled correctly and is in good working order. These tests
should be done after making changes to the probe station configuration but before
cooling down the system. The goal is to find small problems before investing the time
and resources into cooling the system.
These procedures assume that the station has been fully assembled as described in
section 3.6.
3.8.1 Instrumentation
Verification
3.8.1.1 Verifying Voltages
The Model 336 temperature controller is shipped from Lake Shore with user specified
line voltage selected. If the system is being moved, or there is any reason to believe
the voltage may not be appropriate for the installation site, please verify line voltage
settings. The Model 336 temperature controller has a voltage selection with the same
design fuse casing. From the back of the instrument, look at the voltage selection
window to find the voltage setting. Refer to the Model 336 user’s manual if you need
to change the voltage.
3.8.1.2 Verifying Power On
Plug in the power strip (see TABLE 3-1 for power requirements); then, verify that all
the power switch on the back of the controller is in the on position.
3.8.1.3 Verifying the Model 336 Temperature Controller Settings
The control settings for the Model 336 temperature controller are set at Lake Shore;
however, it is good to verify them for safety purposes and to acquaint yourself with
the controller. You can use the tables in section 4.5.5 to assist you in verifying or
resetting it. You can find more information on the controller in its manual.
3.8.2 Temperature
Sensor and Heater Test
Each of the four stages in the Model CRX-6.5K (sample stage, radiation shield stage
and cryocooler first and second stages) has a temperature sensor and heater. Prior to
heating or cooling the system it is important to verify that the sensor, heater, cabling,
and temperature controller are all configured properly. If any of the control loop
wiring is mixed up or a sensor is malfunctioning, it is possible for the system to
overheat and become damaged during warm up. You can check the instrumentation
wiring diagram if your probe station does not perform as these tests suggest
(FIGURE 6-8).
Use steps 1 to 4 to prepare the system for the sensor and heater test. Then use
TABLE 3-7 to run the sensor and heater test.
1. With the system stabilized at room temperature, verify that all sensors read
within a few degrees of each other. Only the sensor on the sample stage is
accurately calibrated, so some discrepancy in the readings can be expected.
2. Install the radiation shield lid and vacuum chamber lid.
3. Use the z-axis micrometers to raise all probes up 3 mm to 4 mm above the
sample stage.
Do not allow the temperature on any stage to raise more than 10 K above room
temperature when the system is not being actively cooled. Carefully observe all
temperature readings and be ready to press the ALL OFF button on the Model 336
temperature controller if any stage raises more than 10 K above room temperature.
4. Use TABLE 3-7 to set the sample stage’s sensor and heater.
5. Monitor the temperature reading. Within 10 min the stage should raise
approximately 5 K and begin to stabilize. The heater output for the chosen stage
should then reduce below 100% and begin to stabilize. This demonstrates that
the sensor and heater are functioning properly and that the control loop is closed.
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...