6.3.4.5 Height Adjustment
The range of focus of the microscope is quite short, as small as 34 µm on some
systems. It is common for a microscope that is set up for thick samples to be incapable
of focusing on the sample holder or thin samples.
Follow this procedure if the microscope does not focus on the sample surface:
1. Set the microscope focus knob to the middle of its range.
2. Loosen the nylon thumbscrew on the horizontal boom.
3. While supporting the microscope with one hand to prevent it from falling,
use the
3/bg
in hex driver to loosen the shaft collar (section FIGURE 6-6).
4. Slide the microscope and collar up or down a few millimeters and tighten the
shaft collar and thumbscrew.
5. Attempt to focus again.
6. Continue adjusting the height until the sample can be focused.
6.3.4.6 Vibration in the Image System
The most common reason for a blurred image on a properly focused microscope is
vibration. Vibration can come from many sources including the CCR itself, vacuum
pumps, motors, and fans running in the vicinity of the probe station.
Vibrations can also affect the microscope directly. Make sure the electrical wires and
fiber optic bundle going to the microscope are not stretched tight and do not pass
over the vacuum line or any other vibration source. Isolate the light source from the
probe station to verify its fan is not causing the blurred image. Finally, verify that the
nylon set screw is tight on the shaft.
Vibration from vacuum pumps:
vacuum pumps used with the probe station are a likely
candidate for vibration because they are in close proximity to the probe station and
are connected through stainless steel lines that transfer vibration. Vacuum pumps
and lines can be easily identified as the source of vibration by observing the
microscope image when the pumps are off and lines disconnected. If a pump or line is
identified as a problem and it cannot be disconnected during operation, the
PS-PLVI-40 pump line vibration isolation option is recommended.
FIGURE 6-6
Microscope focus adjustments
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...