background image

Instructions, Gaussmeter, Cable, and 
HALLCAL.EXE  

 

 

 

 

Lake Shore Cryotronics, 575 McCorkle Blvd. Westerville, OH 4308

www.lakeshore.com

  

Phone (614) 891-2243 

– Fax (614) 891-1362 – 

[email protected]

 

 

[email protected]

  

CAUTION: 

This Hall sensor is sensitive to electrostatic discharge (ESD). Use ESD precautionary 

procedures when handling, or making mechanical or electrical connections to this device in order to 
avoid performance degradation or loss of functionality. 

1.0 

GENERAL 

This document provides operation instructions for operating a Hall sensor with a Lake Shore 420, 421, 
450, or 460 gaussmeter. Using a Hall sensor with a Lake Shore gaussmeter is described in Paragraph 
1.1. MCBL-6 and MCBL-20 programmable cables details are provided in Paragraph 1.2. Finally, setup 
and operation of the Hallcal.exe software is described in Paragraph 1.3. 
 
Instructions in this chapter are intended for the low imped

ance “bulk” Hall sensors. The instructions do 

not

 

apply to HGT-2100, HGT-2010, and HGA-2010 (which are intended for use with the Model 410 
gaussmeter). The impedance and control current requirements of these three Hall sensors are not 
compatible with the Lake Shore gaussmeters that use the MCBL cables. 

1.1 

USING A HALL SENSOR WITH A LAKE SHORE GAUSSMETER 

To hook up a Hall sensor, you must use a Lake Shore MCBL-6 or MCBL-20 cable assembly (Paragraph 
1.2). The MCBL-6 cable is 2 m (6.5 ft) long with a DA-15 connector on one end and four leads on the 
other. The MCBL-20 cable is 6 m (20 ft) long. The Hall sensor is a 4-lead device. The 4 leads are labeled 
+Ic (Red), 

–Ic (Black or Green), +VH (Blue), and –VH (Yellow), corresponding to the 4 leads on all the Hall 

sensors. 

 

The typical Lake Shore 421, 450, or 460 gaussmeter has an input impedance of 420 

Ω. Therefore, the 

actual sensitivity at the gaussmeter input will be less than the value given with the Hall sensor due to drop 
in the leads and cable. This fact is important because a sensitivity value is supposed to be loaded into the 
cable PROM to set calibration. We recommend that the customer always check accuracy against a 
reference field rather than use the sensitivity value sent with the bare Hall generator. Because Lake Shore 
has no control of the conditions beyond the cable, the customer must accept responsibility for accuracy 
and compatibility. 

 

Finally, Manganin wire is not usually acceptable for gaussmeter connections. The resistance of Manganin 
wire is often too high. Hall generators are normally connected using twisted pairs of copper wire such as 
34 gauge, Teflon insulated. There are two reasons for this: 

 

1.  The gaussmeter current source is limited in compliance voltage. The gaussmeter should not drive 

a load (Hall sensor, wires in cryostat, and probe cable) greater than 50 

Ω. In fact, for best 

performance, the load should be less than 30 

Ω. 

 

2.  Because the typical Lake Shore gaussmeter input impedance is 420 

Ω, there is a voltage drop due to 

lead resistance in series with the gaussmeter input. The Lake Shore Hall sensor sensitivity given on 
the data sheet is basically with no lead resistance. See Figure 1-1. 

 

 

 

Impedance2.cvx 

Figure 1-1. Typical Hall Sensor Input Impedance 

Gaussmeter

Input

420

Sensitivity at Gaussmeter input is

reduced by the lead/input voltage divider

R

cable

R

cable

R

cust

R

cust

The gaussmeter needs input sensitivity

between 0.5 to 1.5 mV/kG (HST) or

5.0 and 15 mV/kG (HSE) at its input

V

H

R

cable

= Lake Shore Model

MCBL Cable Assembly

R

cust

= Customer

Supplied Leads

Hall Generator

Open Circuit

Sensitivity

Отзывы: