36
13.7 Setting Offset
Figure 31: Setting Menu
Setting Offset
This setting allows you to offset the displayed temperature from the measured temperature.
14. Troubleshooting
For troubleshooting, use the following:
▪
Alarm status screen
▪
Wiring diagram
▪
Plumbing Diagram
▪
Troubleshooting table (below)
Issue
Possible Cause
Corrective Measures
Clearance By
Unit does not start Power not applied. Electrical
connection not correct or no
mains connection
Check power supply and ensure proper
voltage in the line. Check connection,
insert mains plug.
Operator
Coolant level too low. Alarm
for low coolant will be active
Check coolant level and top off, if
necessary. Refer to Adding Coolant
Operator
Main switch not turned on
Turn main switch on
Operator
Unit running but
cooling capacity is
too low
Buckled or pinched coolant
hoses
Install the hoses with a larger radius to
avoid sharp bends.
Operator
Improperly placed unit
Required clearance with the wall of the
unit. Refer to
Operator
Blocked Heat sink
Clean Heat sink. Refer to
Cleaning the
Heat sink.
Operator
Coolant level too low
Check coolant level and top up, if
necessary. Refer t
Operator
Disconnected coolant hoses Connect the coolant hoses. Refer to
Operator
Dirty coolant filter
Clean or replace filter. Refer to
Operator
Содержание Nextreme NRC400
Страница 1: ...Nextreme Thermoelectric Chiller NRC400 Specification and User Manual ...
Страница 13: ...13 6 Component Locations Figure 2 Isometric Views of Unit Figure 3 Front View Figure 4 Rear View ...
Страница 24: ...24 Plumbing Refrigeration Diagram Figure 10 Plumbing and Refrigeration Diagram ...
Страница 27: ...27 11 2 Pump Performance Figure 14 Pump Performance with Cooling Fluids Water 30 EG W NRC 400 T0 00 PC2 ...
Страница 28: ...28 12 Chiller Dimensions Figure 15 NRC400 Chiller Overall Dimensions ...