INSTRUCTION MANUAL 

 

 

 

 

 

 

 

 
 

 

7000smz 

 

5000mz 

 

Vibrating Blade Tissue Slicers 

 

 

 

 
 

 
 

 

February 2010 

DCN 1310 

Document Ref: 7000-v1.7 Eng 

Содержание CAMPDEN 5000mz

Страница 1: ...www campdeninstruments com INSTRUCTION MANUAL 7000smz 5000mz Vibrating Blade Tissue Slicers February 2010 DCN 1310 Document Ref 7000 v1 7 Eng...

Страница 2: ...lade with the axis of oscillation 6 10 Settings 7 Options 7 1 Light source 7 2 Magnifier 7 3 Microscope 8 Maintenance and Service 9 Returns Decontamination Form 10 Technical support 11 Packing List 12...

Страница 3: ...ion frequencies can lead to excessive vibration and unstable tissue fluid The 7000smz and 5000mz instruments benefit in this area by being built on a heavy rigid cast base giving exceptional immunity...

Страница 4: ...r the uncut tissue has been lowered away from the blade eliminating the possibility of the blade being dragged back across the uncut tissue causing damage Dynamic feedback on the oscillation and advan...

Страница 5: ...s an optional extra The tissue bath is also located and held in place within the outer bath by magnets Inner and outer tissue bath background with standard tissue platen front left and adjustable tiss...

Страница 6: ...A lower power 2x magnifying lens assembly can be fitted as an alternative to the microscope 7000smz fitted with magnifying glass and cold light source...

Страница 7: ...cute cutting edge Because of the relative softness of stainless steel for optimum performance it is recommended that stainless steel blades should be used once only or at a maximum changed every day 7...

Страница 8: ...nsed with clean water on a regular basis The specimen platen and the inner stainless steel bath as shown in the illustration below are autoclavable The specimen holder is located by a circular magnet...

Страница 9: ...be set for your particular voltage supply The voltage is set by prising out the fuse holder drawer and re inserting it such that the voltage legend for your supply is aligned with the mark on the inle...

Страница 10: ...ay Press the MENU key again and the display will change to that shown Use the or keys to scroll to the line SETTINGS Then scroll to the line EDIT USER NAME and select Use the or keys to scroll to the...

Страница 11: ...equency amplitude and section thickness will be remembered under that user name so that when that user comes back to the instrument and selects that user name those settings will also be selected for...

Страница 12: ...m and store the change Frequency settings can be changed in 5Hz steps and amplitude can be changed in 0 25mm steps Note that there are restrictions limiting the amplitudes available for any given freq...

Страница 13: ...ved Once the table has reached its lowest position the display will be as shown The specimen bath is located on the rising table by a convergent double dovetail slideway and clamping screw To remove t...

Страница 14: ...h to a nominal position and the operating screen will change to that shown in the following illustration Note that the SLICE ON OFF key is now illuminated green and the advance speed as shown on the d...

Страница 15: ...ey again to exit This selects the start position such that when the RETURN key is pressed see next page this is the position the blade will return to Press the SLICE ON OFF key to commence the cut The...

Страница 16: ...e start position the section thickness will be highlighted on the display If required the section thickness may be changed by pressing the height button and making any adjustment required via the rota...

Страница 17: ...instrument in the slice window mode without having previously set a slice window via the manual slice procedure the display will indicate NO SLICE PROFILE HAS BEEN CREATED This display will time out a...

Страница 18: ...see section 6 10 If you try to operate the instrument in the automatic mode without having previously set a cutting profile via the manual slice procedure the display will indicate NO SLICE PROFILE HA...

Страница 19: ...w the instructions on screen to CONTINUE or ABORT the procedure Pressing MENU at this stage will allow access to menu functions permitting changes to frequency and amplitude etc Note that the EMERGENC...

Страница 20: ...nt has been halted by use of the Emergency Stop facility it must be reset before operations can continue Twist the red Emergency Stop button clockwise to unlatch it When the button is unlatched the in...

Страница 21: ...s been designed so that the blade may be safely handled without the effective cutting length of the blade edge being exposed to possible damage or injuring the operator After carefully unwrapping the...

Страница 22: ...ht hand screw has a normal right hand thread 5000mz Both left hand and right hand screws have normal right hand threads Once the blade has been fitted it should be aligned to the axis of oscillation i...

Страница 23: ...he blade holder in position once the alignment procedure has been completed See the illustration below Fit the blade holder as shown and tighten the two upper securing screws The lower locking screw s...

Страница 24: ...lied to pass through and reach the set screws 7000smz and 5000mz In order to facilitate alignment of the blade with the axis of travel and so minimise the z axis deflections the instrument uses a demo...

Страница 25: ...appropriate blade using the or keys and press MENU to select Press SLICE to continue The instrument will now re position the blade with respect to the alignment device ready for adjustment The display...

Страница 26: ...screw Conversely the blade holder can be rotated counter clockwise by first loosening the left hand set screw and then tightening the right hand screw Press SLICE to set the blade vibrating again and...

Страница 27: ...keys Press the MENU key to confirm the change Use the or keys to scroll to line EXIT MENU and press the MENU key to get back to the basic menu display 2 When in an automatic slice mode the park positi...

Страница 28: ...rce can be switched on and the light intensity controlled via a combined rotary switch potentiometer on the body of the light source where the flexible arms are mounted The output beams may be focusse...

Страница 29: ...ifier should then be fixed to the top of the instrument cover using the screws provided Do not over tighten the screws This is illustrated below If you also have the optional light source the magnifie...

Страница 30: ...ification are supplied The microscope is attached to the instrument via screws to a dedicated mounting pad on the rear of the instrument this ensures a rigid mount rendering the microscope free from v...

Страница 31: ...ll not be accepted for service or repair unless the unit has been adequately and properly packaged Additionally instruments will not be accepted without prior authorisation and have been certified as...

Страница 32: ...e Sources YES NO IF YES DO NOT RETURN 3 Chemical Hazard mercury salts acids bases etc YES NO 4 Rabies BSE CJD nCJD etc YES NO 5 Any Other Hazard PLEASE SPECIFY YES NO C Does your laboratory contain an...

Страница 33: ...itnessed This information may be readily accessed from the basic menu screen as follows Use the or keys to scroll to the line ABOUT and press the MENU key The screen will now display the parameters re...

Страница 34: ...ibrating head has been obstructed restart the instrument and carefully drive the blade away from the obstruction If the error occurs repeatedly during normal use contact Technical Support for advice C...

Страница 35: ...1 off Opti Cal optical Calibration unit 1 off optional extra with 5000mz Stainless steel blades 10 off 7550 1 SS Ceramic blades 2 off 7550 1 C Cross head Philips PoziDriv screwdriver 1 off Hexagonal...

Страница 36: ...7000 3 3 Specimen holders Replacement specimen holder standard 7000 4 1 Replacement adjustable specimen holder 7000 4 2 Blade holders 5000mz Blade holder 7000 5 1 Also fits 7000smz prior to serial nu...

Страница 37: ...n frequency 7000smz Minimum 50 Hz Maximum 120 Hz amplitude dependent 5000mz Minimum 50 Hz Maximum 100 Hz amplitude dependent Frequency step size 5 Hz Blade oscillation amplitude 7000smz Minimum 0 5 mm...

Страница 38: ...rews for future use If it is necessary to return the instrument for service or repair please repack the instrument in the original packing The packing procedure is a reversal of the above procedure en...