SimNewB
™
ENGLISH
Directions for Use
www.laerdal.com
SimNewB.qxp:SimNewB DFU EN 3/25/08 4:22 PM Page 1
Страница 1: ...SimNewB ENGLISH Directions for Use www laerdal com SimNewB qxp SimNewB DFU EN 3 25 08 4 22 PM Page 1...
Страница 2: ...SimNewB qxp SimNewB DFU EN 3 25 08 4 22 PM Page 2...
Страница 3: ...ON General handling 11 Preset Patient State Levels 12 Airway simulation features 13 Breathing simulation features 15 Circulation simulation features 17 Drug IV administration 19 Manikin preparation be...
Страница 4: ...his device must accept any interference received including interfe rence that may cause undesired operation NOTE This equipment has been tested and found to comply with the limits for a Class B digita...
Страница 5: ...ilateral Simulated blood flashback upon cannulation Sounds Vocal Grunt breathing crying hiccups and others Lung Normal Stridor pneumonia and others Heart Normal systolic murmur and others Remote Contr...
Страница 6: ...Bag andTubing Kit 220 00750 Blanket 220 00650 Baby pants 220 00450 Hat 200 03050 Eye Kit 250 21050 Airway Lubricant 300 00750 Red Simulated Blood 4 oz 240 00250 Umbi Simulated Blood 4 oz 276 15550 Liq...
Страница 7: ...from the manikin blood pressure connector H Battery case Holds 6 replaceable C cell batteries Overview of Remote Control I LCD display Displays status and operation information J Quick Execution keys...
Страница 8: ...20 capacity remains a Battery low message will be displayed by the Remote Control When the remaining capacity in the Control Unit battery is less than 10 a Replace battery warning will by given on th...
Страница 9: ...r supply cable into power source 5 Connect the Air CO2 tube to the Output to Manikin connector 2 Optional this is not required for operation 6 Connect the CO2 regulator to a suitable CO2 tank 7 Connec...
Страница 10: ...Pressure Cuff s connector to the Manikin Blood Pressure Cuff Connector The BP cuff may be wrapped around the right upper arm 7 Connect the empty IV bag set to the Umbilical Reservoir Overflow tube No...
Страница 11: ...ing any selections Pressing the on off key R will switch off the remote Control Language One of several languages can be selected Language selection affects the text on the Remote Control display Firs...
Страница 12: ...ber of mi nutes A warning will be given one minute before the unit turns off Factory setting is 15 minutes Com Channel To be able to operate several SimNewB orVitalSim systems by wireless remote contr...
Страница 13: ...nd motions The muscle tone of the SimNewB s limbs can be set to repre sent either a limp poor muscle tone or a normal newborn The muscle tone impression can be enhanced with spontane ous vigorous moti...
Страница 14: ...ygroma polycystic kidneys cleft lip palate 2 Preset Patient State Levels The NewSimB Neonatal Simulator system has six predefined patient state levels L0 L5 For each level seven clinical parame ters a...
Страница 15: ...stigating proper procedures for removal of meconium The head can be turned to the side as normal Clearing the lower airway One can dry simulate removal of meconium from the SimNewB manikin s mouth an...
Страница 16: ...disappear An ET tube may be secured by means suitable for a neonatal baby Tape residues should be cleaned off the skin with soap and water To simulate airway obstructions or very low lung compliance...
Страница 17: ...Specifications section page 36 When connected to a CO2 source the SimNewB manikin can expire CO2 This function can be turned on and off from the Remote Control or the PC Software Central cyanosis To...
Страница 18: ...g the stomach To vent out accumulating air from the stomach an orogastric tube size 8 FR recommended can be in place simultaneously with mask ventilation Breathing sounds audible by auscultation Breat...
Страница 19: ...to the simulation team members on the provided moni tor 5 Circulation Heart rate and Pulse The SimNewB manikin has a heartbeat and palpatable umbilical and right brachial pulses The palpatable pulses...
Страница 20: ...sounds Heart sounds can be listened to with a stethoscope at the left side of the chest The volume can be adjusted from level 1 to 9 Level 7 is default The available heart sounds are listed in the Det...
Страница 21: ...ded to the reservoir so that simulati on team members can draw simulated blood when testing the catheter for proper insertion depth Intraosseous access Intraosseous access can be established by cannul...
Страница 22: ...so will result in a moving umbilicus when pulse is on To remove the umbilical cord just pull it out directly Changing eye pupils 1 Open the eyelids wide take care not to rip the faceskin 2 Using the s...
Страница 23: ...and wipe it clean and let it dry Suction fluids from the reservoir overflow connection Add clear or soapy water and suction until the reservoir is clean Emptying leg for fluid If fluid is present in t...
Страница 24: ...ed into six functional areas and a Quick Execution function line 1 Log functional area shows the log clock log status and if the log has been saved the current log s file name 2 Radio reception indica...
Страница 25: ...nus Pressing and holding the shift key gives access to a se cond set of quick execution functions 2 Numbered parameters can be altered directly with the nu meric keys 3 Pressing the OK key will take y...
Страница 26: ...manikin s CO2 output on off Selected Lung Closures quick execution functions SeeTable 1for the interpretation of the lung closu res symbols Set Airway Parameters menu When the Breathing functional are...
Страница 27: ...et Airway Parameters menu pressing the icon quick execution key opens the Set Lung Sound menu Figure 8 Pressing C no change or OK activate change re turns you to the Set Airway Parameters menu Use the...
Страница 28: ...right of the selection field indicates that there are more selections than presented on the screen You can also en ter the numeric identifier directly with the numeric keys The heart rate pulse if act...
Страница 29: ...BP quick execution keys When Systolic is selected as in Figure 15 only systolic will change but diastolic will be kept at least 10 mmHg below systolic When Diastolic is selected only diastolic will ch...
Страница 30: ...ys The arrow at the right of the selection field indica tes that there are more selections than presented on the scre en Press OK to activate selections The last three vocal sounds used become availab...
Страница 31: ...i bed below Use the Limp Tone Motion Seizure quick execution keys to select muscle tone state The Resp quick execution key toggles the manikin s chest rise on off When breathing rate is 0 there will b...
Страница 32: ...Log The Save quick execution function pauses the log and opens the Save Log field The log is given a default name by the SimNewB system LOGxx where xx is a sequential number This name can be modified...
Страница 33: ...OK The scenario will start with the clock set to 00 00 and a new log will begin The current log will be erased unless it is saved befo re the scenario is started When running a scenario the top line o...
Страница 34: ...tallation Insert the software CD in the CD ROM drive and wait for the installation program to start If it does not start automatically run the Setup exe program located on the CD Follow the instructio...
Страница 35: ...unit Log and Scenario Control This utility is used to load scenarios into the SimNewB system and to retrieve logs that have been saved in the SimNewB sys tem Connect the SimNewB Control Unit to the P...
Страница 36: ...ion 915 5 916 4 MHz Operation range 10 m 30 ft max BP cuff pressure input 0 300 mmHg Sinus Sinus w Unifocal PVCs Sinus w Couplet PVCs Sinus w Multifocal PVCs Sinus w PACs Sinus w PJCs AtrialTachycardi...
Страница 37: ...n C Upright QRS with ST elevation D BBB Broad R wave withT inversion E Biphasic QRS withT inversion F QS with ST elevation G BBB Broad S wave with uprightT Ventricular QRS types for Idioventricular V...
Страница 38: ...kles Pneumonia Wheeze Rhonchi Stridor Wheezes Content Hiccups Grunting Strong Cry Weak Cry Cough Scream Blood Pressure Pressure range 0 300 mmHg Accuracy 4 mmHg Calibration Pressure sensor must be cal...
Страница 39: ...mote Control not working 1 Check the batteries for both units Check that the paper shipping strip has been removed from between batteries 2 Connect the phone cable from the Remote Control to the Contr...
Страница 40: ...2008 Laerdal Medical Corporation All rights reserved Printed in USA Part no 1005452 RevA SimNewB qxp SimNewB DFU EN 3 25 08 4 24 PM Page 1...