laerdal 801-00233 Скачать руководство пользователя страница 1

CPRmeter 2

User Guide

EN

www.laerdal.com

Содержание 801-00233

Страница 1: ...CPRmeter 2 User Guide EN www laerdal com...

Страница 2: ...d 12 CPRmeter 2 Placement 13 Feedback Display Overview 15 Compression Feedback 16 Depth 16 On Soft Surface 17 Rate 18 Compression Counter 18 Inactivity 19 Debriefing 20 Battery Indicator 22 Maintenanc...

Страница 3: ...number of compressions in a series and provides notification of lack of expected CPR activity If in doubt about the appropriateness for use perform CPR without using the CPRmeter 2 Indication for Use...

Страница 4: ...6 7 Items Included Items Included A A A X 2 Protective Sleeve Patient Adhesive AAA Batteries CPRmeter 2...

Страница 5: ...8 9 Overview Overview On Off Status Light Compression Area Battery Hatch Display Area Placement guide for adhesive Placement guide for adhesive Membrane Port...

Страница 6: ...se Insert Batteries Before Use Apply Patient Adhesive 1 1 x 2 2 4 3 1 3 2 Cautions Ensure Patient Adhesives are within their expiration date Adhesives should be removed from the device and disposed of...

Страница 7: ...rface Remove clothing from the patient s chest Turn On Place CPRmeter 2 If needed wipe fluids off the chest Ensure to use the heel of your hand and apply pressure to the light grey area Begin CPR Prov...

Страница 8: ...pport after the event Warnings Do not use the device in conjunction with any mechanical or automated compression device Do not use the device on top of defibrillation pads unless the manufacturer of t...

Страница 9: ...ensuring that for each chest compression the area below the compression depth target lights up Warning When performing CPR on a patient lying on a mattress a backboard must be used to limit the amoun...

Страница 10: ...r digits flash solid white for every tenth compression Without a compression the counter is reset after 3 seconds Inactivity After 3 seconds the CPRmeter 2 displays an inactivity timer which counts th...

Страница 11: ...Rmeter 2 is turned off When turned on again the statistics from the last stored CPR event can be reviewed When the CPRmeter 2 is used in a new CPR event the Q CPR Quick Review will display the current...

Страница 12: ...en the CPRmeter 2 is being turned on 23 Note If the remaining battery power during use becomes too low to sustain further operation the low battery icon is shown for 10 seconds and then the CPRmeter 2...

Страница 13: ...erior surfaces and allow to sit for a minimum of 12 minutes An alternative disinfection agent is isopropyl alcohol 70 solution If necessary respray to account for evaporation of isopropyl alcohol 2 Wi...

Страница 14: ...C to 40 C 32 F to 104 F Sealing Meets ISO IEC 60529 class IP55 Product temperature can reach 60 C 140 F Relative Humidity Transport 5 to 95 Storage 5 to 75 Operation 5 to 95 Atmospheric Pressure Atm...

Страница 15: ...Directive 1999 5 EC Compliance with applicable U S and Canadian safety standards has been certified by Canadian Standards Association These CPRmeter 2 patient adhesives are disposable and are for sin...

Страница 16: ...t necessary to remove the device from the patient Do not practice by using the CPRmeter 2 on a person It may be used with a training manikin or simply on a compliant surface for practice Properly perf...

Страница 17: ...ion Statement and Industry Canada Statements This device complies with part 15 of the FCC rules and RSS 210 of the Industry Canada rules Operation is subject to the following two conditions 1 This dev...

Страница 18: ...dings used for domestic purposes Electromagnetic Immunity Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge ESD IEC 61000 4 2 6 kV contac...

Страница 19: ...nt Electromagnetic Emissions Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2...

Страница 20: ...s applies to the model 801 00233 CPRmeter 2 This information is subject to change The CPRmeter with Q CPR is protected by U S and International registered patents The design of CPRmeter and its feedba...

Страница 21: ...Rev A Laerdal is a registered trademark of Laerdal Medical AS 2015 Laerdal Medical AS All rights reserved P O Box 377 Tanke Svilandsgate 30 4002 Stavanger Norway T 47 51511700 Printed in Norway www la...