8 MAINTENANCE AND SERVICE
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
137
Pump Fan Filter
–
Tower Configuration
TITAN VERSA Leak Components
Part Number
Description
CM112498/L-7/4
Calibrated Leak
PFP0488E1
Temperature Board
LSLC0001
Calibrated Leak Standard Calibration, Single Point
LSLC0003
Sniffer Calibrated Leak Standard Calibration, Single Point
Recalibration
NOTICE:
Leak Standard Calibration Frequency
LACO recommends that the internal calibrated leak standard be calibrated at
least every two years.
Sniff calibrated leaks and higher leak rate external leaks (> 1.0 e-6 mbar*L/sec)
should be calibrated yearly, or more frequently if the depletion rates are large.
•
Depletion rate is indicated on the calibrated leak identification label.
•
Return leak standards to LACO Technologies calibration lab for recalibration
.