6 OPERATION
96
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Exponent = Integer value (U) - 12
He Compound = Mantissa x 10
Exponent
Examples
•
U = 3.91 V -> He Compound = 10 x (3.91-3)) x 10
(3-12)
= 9 10 10
-9
•
U = 8.25 V -> He Compound = 10 x (8.25-8)) x 10
(8-12)
= 2.50 10
-4
Inlet Pressure
The "Inlet Pressure" output corresponds with the inlet pressure source:
Formula
U = Voltage measured (V) on analog output
Inlet pressure = 10
(U-
5.5)
hPa