LABORIE Urocap IV Owner’s Manual UC4-UM01
45
19
Wait for the data to be transferred to the
i-LIST™ Office
Reporter
application.
20
In the fields on the left side of the screen, enter information to include in the printed report.
21
Click the printer icon to print the report.
22
Empty the beaker. It can be thoroughly washed and reused.
Table 15 – Performing a Uroflow Study using Automatic Flow Detection
7.2
MANUAL METHOD
STEP
1
Gather the supplies needed for a Uroflow test. (Urocap™ IV, beaker, commode chair, funnel, and
straight catheter).
2
Plug the power cord into an electrical outlet.
3
Place the funnel on the plastic frame of the commode chair.
4
Carefully place the Urocap™ IV on the floor or on an approved Uroflow transducer stand.
5
Gently place a graduated beaker on top of the Urocap™ IV.
6
Place the commode chair and funnel over the Urocap™ IV and beaker. Make sure the beaker and the
funnel are aligned but not touching.
7
Turn on the computer and then turn on the printer.
8
Invite the patient to get ready to take the test.
9
Double-click the Urocap™ IV shortcut icon on the
computer desktop.
10
Make sure the Urocap™ IV is connected.
11
Click the Uroflow button on the control panel.
12
Click Set Zeroes. Click Run to start the test.
13
CAUTION! DO NOT TOUCH the beaker during voiding.
14
Instruct the patient to void; provide privacy if possible.
15
When voiding is complete click Stop.
16
Straight catheterize patient to obtain post void residual measurement.
Remove the catheter.
17
Click Info and Voiding Summary.
Enter the post-void residual volume in the PVR box.