114
Nexam Pro User Manual
Chapter 9
Click the
Catheter
button to display the Catheter selection window. Select
your catheter from the list; for example, the disposable water catheter
LABORIE G-84481
or the
LABORIE G-84482
.
Radial catheter mode
In the
Edit investigation protocol
window, you can select the catheter mode:
staggered/helix (default selected) or radial mode. When radial mode is
selected, on the
Test selection
tab, the resting pressure test, the squeeze test
and the push test are combined in one sequence. When the rest, squeeze or
push test is ended, the next test in the combined sequence will be activated.
When the cm marker is inserted during the sub sequence, the sub sequence
will start from the beginning. If you want to change the catheter insertion depth
during the investigation, you can use the
Go to x cm
button (puller) or the
Mark x cm
button (manual pull).
When right-clicking this button, the x cm is reset to initial value (when a puller
is used), or the insertion depth is increased with 1 cm (when manual pull is
used).
When radial mode is selected, the software will ask if RAIR is present or
absent when inserting the RAIR end marker. In the results, this is shown in the
marker name and it is added to the report. The balloon volume is shown in the
RAIR marker hint.
When the investigation has been performed, and the analysis program is
started, the results are calculated on all channels.
9.2.3
Select the Anorectal Manometry Tests
In the protocol settings, you can select which tests can be performed during
the investigation, as well as adapt the order of the selected tests. Click the
More
button and then select the
Test Selection
tab. The window shown in
Figure 9.4 is displayed.