Cleaning procedure:

Place the objectives and/or eyepieces on a

 

dust-free surface (e.g. fresh aluminum foil). All other optical components to 

be cleaned should be as accessible as possible.

1.  Blow all loose dust particles away with a dust blower.
2.  Remove all water-soluble dirt with distilled water. If this is unsuccessful repeat using a solution of diluted hand 

soap liquid. Remove any remaining residue with a dry cotton swab.

3.  To remove oil, use a solution of diluted hand-soap liquid initially. If this does not produce a satisfactory result, 

repeat the cleaning using a solvent (Optical Cleaning Solution 85% petroleum ether and 15% isopropanol).

4.  Grease must always be removed using a solvent.
5. Cleaning is achieved by using a spiral motion from the center to the rim. Never wipe using zig-zag movements as 

this will only spread the dirt. With larger optical surfaces (e.g. tube lenses) the spiral motion starts initially at the rim 
before moving to the middle and is only then followed by a center to rim cleaning motion. Normally several spiral 
wipes are recommended.
We recommend pure, volatile petroleum ether or Optical Cleaning Solution as explained in point 3 above.

Wipe using a spiral movement. Do not use a zig-zag motion!

2. 

Cleaning of painted surfaces :

Avoid the use of any organic solvent ( e.g. thinner, xylene, ether, alcohol etc.) for cleaning of painted surfaces of 
the instrument. Painted surfaces can be cleaned with a very lightly moistened micro fiber cloth. Loose dust and 
other dirt can be removed using a brush of soft hair used exclusively for this purpose.

3.

Cleaning of plastic surfaces:

Sigma microscope frame is made up of special grade plastic which can be cleaned with mild soap solution. Do not 
use acetone for cleaning stage condenser lens.

Caution: 
Do not use aggressive organic solvent such as acetone for cleaning painted surfaces and plastic parts of 
the microscope. 

4. 

Never attempt to dismantle :

Never attempt to dismantle the instrument so as to avoid the possibility of impairing its operational efficiency and 
accuracy.

5. 

Periodical checking :

To  maintain  the  performance  of  the  instrument,  we  recommend  customers  have  their  microscopes  serviced 

periodically by a factory authorized dealer/rep. For details, contact your nearest dealer or Labo America’s  main 
office in California.

zig-zag motion (X)

spiral motion (   )

!

4

Sigma

Содержание Sigma

Страница 1: ...ure proper use of this instrument as well as to avoid injury while operating instrument understanding this manual completely before use is highly recommended LABOMED User Manual Educational Microscopy...

Страница 2: ......

Страница 3: ...IFICATIONS 6 10 DETAILED OBSERVATION PROCEDURE 13 14 9 SUMMARY OF BRIGHTFIELD OBSERVATION PROCEDURE 12 7 INITIAL SETUP 10 6 OPTIONAL ACCESSORIES 9 8 5 STANDARD COMPONENTS 7 4 SYSTEM DIAGRAM 6 3 UNPACK...

Страница 4: ...e free operation The sturdy new stylish design provides comfort as well as stability The high powered objectives are spring loaded to prevent accidental damage to specimen slides The quadruple nosepie...

Страница 5: ...charging adapter provided by LABOMED If the proper charging adapter is not used product safety performance cannot be warranted 5 When installing the microscope route the charging adapter away from th...

Страница 6: ...safe operational life with the least amount of maintenance required In general routine maintenance is limited to keeping the microscope working parts lubricated and optics clean Always cover the micro...

Страница 7: ...ecommend pure volatile petroleum ether or Optical Cleaning Solution as explained in point 3 above Wipe using a spiral movement Do not use a zig zag motion 2 Cleaning of painted surfaces Avoid the use...

Страница 8: ...UNPACKING YOUR MICROSCOPE 3 5 Power Adapter Battery charger Microscope Mechanical stage optional Slides optional...

Страница 9: ...6 SYSTEM DIAGRAM 4 Sigma Integrated eyepiece with eye guard Objectives Stage clips Coarse focusing knob Fine focusing knob...

Страница 10: ...STANDARD COMPONENTS 5 7 After opening the package make sure that the correct units for the selected set are present LED Adapter Integrated eyepiece Objectives Stage clips...

Страница 11: ...wards in direction of arrows as shown in figure 3 2 Align the indexing pins A and B to the index holes 1 and 2 provided on the stage plate Align A with 1 and B with 2 as shown in figure 4 3 Fix the st...

Страница 12: ...er positioning your specimen slide return the bow shaped lever 4 gently by slowly releasing lever handle 3 3 Use X axis movement control knob 2 and Y axis movement control knob 1 for horizontal and ve...

Страница 13: ...ovided at rear of the microscope See figure 6 Fig 6 Objectives are factory set Objectives are par centered and par focalised during assembly phase All objectives have been secured for a tight fit to p...

Страница 14: ...4x objective 1 Raise the stage to the highest position by rotating the coarse focus knob 7 clock wise See figure 8 2 Use allen key 3mm 2 to unlock the screw 1 from the stage to un secure LED assembly...

Страница 15: ...te the knob towards I ON Place the specimen on the stage Engage the 10X objective in the light path Bring the specimen in focus Engage the objective to be used in the light path and bring the specimen...

Страница 16: ...e 11 1 Rotate the coarse adjustment knob 1 anticlockwise to fully lower the stage 2 Press the stage clips 2 one by one with finger to lift them from front Place the specimen by sliding the specimen gl...

Страница 17: ...is obtained an image is detected rotate the fine adjustment knob 2 for fine focusing Working Distance WD The WD refers to the distance between each objective and the specimen when precise focus of the...

Страница 18: ...4 Part of image is defocused or image The objective is not properly engaged Engage the objective into position looks like it s flowing in the light path until it clicks The specimen is not set properl...

Страница 19: ...imen 6 Dimensions 267 0mm L x 179 6mm W x 368 0mm H 7 Electrical Battery 3 7V 500mAH Charging time up to 3 hours with totally consumed battery Back up time up to 8 hours 8 Operating environment Indoor...

Страница 20: ...009 Labo America Inc 2124000 990A 12 2009 LABOMED and Sigma are registered trademarks of Labo America Inc With a policy of continuous development Labo America Inc reserves the right to change design a...

Отзывы: