Pressing the Start/Stop Button

 



Run and Time are illuminated. The timer will count down from the set time value.

NOTE:

 You cannot adjust the set time during shaking.

99.4

Stop

Encoder

Power

Run

Time [min]

RPM [x10]

Temp [°C]

Start

RMP Adjusting During a Run

 



Push the Encoder knob until the RPM light is illuminated.

 



Rotate the encoder right (+) or left (-) until the new value is set. While adjusting the RPM, the light will blink. 

 



When you stop rotating the encoder knob, the RPM light will stop blinking after 2 sec.

l l

 l l l l l l l

100

Stop

Encoder

Power

Run

Time [min]

RPM [x10]

Temp [°C]

Start

Temperature Adjusting During a Run

 



Push the Encoder Knob until the RPM light is illuminated.

 



Rotate the encoder right (+) or left (-) until the new value is set. While adjusting the RPM, the light will pulse. 

 



When you stop rotating the encoder knob, the RPM light will stop blinking after 2 sec.

l l

 l l l l l l l

37.0

Stop

Encoder

Power

Run

Time [min]

RPM [x10]

Temp [°C]

Start

 



When the set time expires or if you press the Start/Stop button, End will be displayed and the Run light will blink. When the unit 
stops shaking, the last set values for Time, RPM, and Temp will be saved in memory.

7.3. Shaking Without Heating

If you want to shake samples without heating, set Temp to OFF.

7.4. Heating Without Shaking

If you want to heat samples without shaking, set RPM to OFF.

NOTE:

 The shaker motor drives the fan, which circulates warm air throughout the chamber aiding in temperature uniformity. If 

the RPM is set to OFF, the temperature control will not be as precise due to lack of airflow over the heating element. Maximum set 
temperature in this program is 70°C.

Содержание VorTemp 56

Страница 1: ...Labnet VorTemp 56 Shaking Incubator Instruction Manual Catalog Numbers S2056A S2056A 230V...

Страница 2: ...orm 3 5 0 Safety Information 4 6 0 Installation 4 6 1 Device Placement 4 6 2 Attaching the Power Cord 4 6 3 Start Up Check List 4 7 0 Operating Instructions 5 7 1 Control Panel 5 7 2 Basic Operation 5...

Страница 3: ...Electrical S2056A S2056A 230V 120V 10 50 60Hz 4 8A 230V 10 50 60Hz 2 4A Fuse S2056A S2056A 230V 2 x T 6 3A L 250V 2 x T 3 15A L 250V Environmental Temperature From 5 C to 40 C Relative Humidity Up to...

Страница 4: ...temperature sensor and the control electronics regulate heater function to maintain the set temperature 4 0 Accessories Cat No Description Qty Pk S2056 R 1 5 2 0 mL platform workstation for microcentr...

Страница 5: ...stacked The working temperature range is ambient from 5 C to 40 C Please note that the normal maximum temperature for ordinary microplates is below 60 C The platform is made of stainless steel and is...

Страница 6: ...ss to the incubator control panel and main plug in case of emergency When selecting the right place for device please consider following Put the device on a smooth horizontal and stable place Leave at...

Страница 7: ...ture 6 Encoder Knob Rotate the encoder right or left to modify the Time Temperature or RPM settings Push in the Encoder knob to change between Time Temperature and RPM set values If you rotate the Enc...

Страница 8: ...Temp C Start RPM Illuminated Rotating the encoder right or left sets the rotating speed to the desired value 34 340 rpm 90 900 rpm By turning the RPM setting below 20 the shaking function can be turn...

Страница 9: ...left until the new value is set While adjusting the RPM the light will pulse When you stop rotating the encoder knob the RPM light will stop blinking after 2 sec l l l l l l l l l 37 0 Stop Encoder Po...

Страница 10: ...with the motor regulation PWM regulator pulse generator motor The unit will automatically stop Call for service E 1 2 The motor does not reach the set RPM in 30 sec The unit will automatically stop Ca...

Страница 11: ...chase CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be...

Страница 12: ...SIA PACIFIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 India t 91 124 4604000 Japan t 81 3 3586 1996 Korea t 82 2 796 9500 Singapore t 65 6572 9740 Taiwan t 886 2 2716 033...

Отзывы: