Instruction Manual |
7
Pressing the Start/Stop Button
Run and Time are illuminated. The timer will count down from the set time value.
NOTE:
You cannot adjust the set time during shaking.
99.4
Stop
Encoder
Power
Run
Time [min]
RPM [x10]
Temp [°C]
Start
RMP Adjusting During a Run
Push the Encoder knob until the RPM light is illuminated.
Rotate the encoder right (+) or left (-) until the new value is set. While adjusting the RPM, the light will blink.
When you stop rotating the encoder knob, the RPM light will stop blinking after 2 sec.
l l
l l l l l l l
100
Stop
Encoder
Power
Run
Time [min]
RPM [x10]
Temp [°C]
Start
Temperature Adjusting During a Run
Push the Encoder Knob until the RPM light is illuminated.
Rotate the encoder right (+) or left (-) until the new value is set. While adjusting the RPM, the light will pulse.
When you stop rotating the encoder knob, the RPM light will stop blinking after 2 sec.
l l
l l l l l l l
37.0
Stop
Encoder
Power
Run
Time [min]
RPM [x10]
Temp [°C]
Start
When the set time expires or if you press the Start/Stop button, End will be displayed and the Run light will blink. When the unit
stops shaking, the last set values for Time, RPM, and Temp will be saved in memory.
7.3. Shaking Without Heating
If you want to shake samples without heating, set Temp to OFF.
7.4. Heating Without Shaking
If you want to heat samples without shaking, set RPM to OFF.
NOTE:
The shaker motor drives the fan, which circulates warm air throughout the chamber aiding in temperature uniformity. If
the RPM is set to OFF, the temperature control will not be as precise due to lack of airflow over the heating element. Maximum set
temperature in this program is 70°C.