Instruction Manual |
5
5.2. Basic Operation
Press the POWER switch on control panel. The Labnet Orbit™ Digital Shaker automatically detects supply frequency of 50 or 60Hz,
and displays F50 or F60.
F50
STOP
Encoder
POWER
START
RUN
TIME
RPM
Push
Then the display switches to show the set time and illuminates the Time light.
99.5
STOP
Encoder
POWER
START
RUN
TIME
RPM
Push
Time signal light is on. By rotating the encoder knob right (+) or left (-) the set time can be selected from 30 sec. to 99 min. 50 sec.:
•
99.5 = 99 min. 50 sec.
•
9.59 = 9 min. 59 sec.
•
0.30 = 30 sec.
If you want to set the timer to Hold, turn the knob left or right until “HLd” is displayed.
You can reach the “HLd” function by setting to under 0.30 or above 99.5.
HLd
STOP
Encoder
POWER
START
RUN
TIME
RPM
Push
Push the encoder knob to switch modes between time set mode and RPM set mode. RPM signal light is on. By rotating the encoder
knob right (+) or left (-) the desired RPM can be selected:
•
34 = 340 rpm
•
120 = 1,200 rpm
120
STOP
Encoder
POWER
START
RUN
TIME
RPM
Push