18
| LABNET MULTIGENE OPTIMAX THERMAL CYCLER
12.2 Lid Temperature
Users can change the temperature of lid of this unit by selecting Lid Temperature screen under SYSTEM screen.
NOTE:
Remember to
press F1:ACCEPT to store your settings. Two temperature segments: 60°C to 65°C and 100°C to 115°C are available.
SYSTEM>LID TEMPERATURE
Lid Temperature Function:
1
8/21/2022
11:00
(1:ON
2:OFF)
Lid Temperature:
105
(60--65 ),(100--115)
C
Press F1 to save your settings
ACCEPT
EXIT
F 1
F 2
F 3
F 4
F 5
12.3 Beeper
Users can turn ON or OFF the sound of this unit by selecting the Beeper screen under SYSTEM screen.
NOTE:
Remember to press
F1:ACCEPT to store your settings.
SYSTEM>BEEPER
8/21/2022
11:00
Beeper Function:
2
(1:ON
2:OFF)
Press F1 to save your settings
ACCEPT
EXIT
F 1
F 2
F 3
F 4
F 5