8.0 Protocol Management
After entering the selected User folder, users can press F1-F4 keys to Run, New (Create), Edit (View), and Delete a selected Protocol as
shown below.
Mike>PROTOCOL
PROTOCOL
8/21/2022
11:00
DATE
TIME
Please select a protocol and press
RUN
NEW
EDIT
DELETE
EXIT
Touchdown
07/25/11
09:10:20
Lab
07/28/11
03:25:30
F 1
F 2
F 3
F 4
F 5
8.1 Running a Protocol
1. Press F1:RUN to run the selected Protocol under the User folder screen. The screen will show as below.
Mike>Lab>RUN
1 Tmp
8/21/2022
11:00
3 Tmp 25 Cyc
2 Tmp 1 Cyc
Temp
94.0 94.0 55.0
Time
5:00
0:30
0:30
72.
0
72.00 4.00
10:00
Mode:1(1-2)
Volume:
50
ul
(5-100ul)
1:Tube Control 2:Block Control
START
EXIT
F 1
F 2
F 3
F 4
F 5
2. Users are required to select a Temperature Control Mode before running the Protocol.
Mode 1 (Tube Control):
Enables system operation with simulated temperature control. If Mode 1 is selected, users are required to
input the sample volume: 5-100 µL. In this mode, the temperature control is based on Block temperature and an algorithm which
estimates the sample temperature based on the block temperature and the sample volume. Recommended for more accurate
temperature control.
Mode 2 (Block Control):
Enables system operation with maximum temperature control based on the actual Block temperature.
Recommended for normal operation.