background image

1. Components and functions

(1) Plunger

Depressing the plunger dispenses the set volume of liquid

(2) Volume adjustment dial

Select position 1 to 5 to set desired volume (see table 1)

(3) Fill lever

Fills tip with liquid when pushed upwards

(4) Locking lever

Secures reservoir tip in place

(5) Reservoir tip

Reservoir for holding liquid to be dispensed. For best
results use Labnet Combi-Syringes.

2. Selecting the volume

Locate the volume to be dispensed in table 1

Set the volume adjustment dial to the position
indicated at the top of the column.

Choose the appropriate reservoir tip from the left
hand column of the chart.

The maximum number of dispensing steps for the
selected tip and volume is located at the bottom of
the column.

  3.   Inserting reservoir tip

  •   

Push the fill lever down until it stops.

  •  

 Push the locking lever upwards.

  • 

  Insert a reservoir tip with slight pressure to lock it in

   place.

  •   

Push the fill lever down slightly and press down on

   the locking lever to secure the tip in place.

  

  Use the volume adjustment dial to set the required

   volume.

Labpette R

Instruction Manual

                                                                      

               Table 1

Dial position

1

2

3

4

5

using   0.05 ml tip

µ

l

µ

l

µ

l

µ

l

µ

l

using   0.50 ml tip

10 

µ

l

20 

µ

l

30 

µ

l

40 

µ

l

50 

µ

l

using   1.25 ml tip

25 

µ

l

50 

µ

l

75 

µ

l

100 

µ

l

125 

µ

l

using   2.50 ml tip

50 

µ

l

100 

µ

l

150 

µ

l

200 

µ

l

250 

µ

l

using   5.00 ml tip

100 

µ

l

200 

µ

l

300 

µ

l

400 

µ

l

500 

µ

l

using 12.50 ml tip

250 

µ

l

500 

µ

l

750 

µ

l

1000 

µ

l

1250 

µ

l

using 25.00 ml tip

500 

µ

l

1000 

µ

l

1500 

µ

l

2000 

µ

l

2500 

µ

l

using 50.00 ml tip

1000 

µ

l

2000 

µ

l

3000 

µ

l

4000 

µ

l

5000 

µ

l

     Steps

  48

  23

  15

  11

  8

Please note :

25 ml und 50 ml tips require the use of an
adapter (adapters are supplied free with
each Labpette R repeating pipettors).

    Attach the adapter by placing it on the tip

as shown in the illustration and turning
clockwise. After the tip has been used,
the adapter may be removed and reused.

    0,05 ml tips should be used with a 10 

µ

l

    pipette tip on the liquid end.

4. Filling the reservoir tip

 Immerse the end of the reservoir tip into the liquid to
 be drawn up.

 Slowly raise the fill lever upwards to draw liquid into
 the tip.

• 

 Dispense the first step with the plunger.

The fill lever must be raised slowly to
prevent negative pressure and air
bubbles in the liquid.

    A small air bubble in the upper part of
   the reservoir will not affect the
   accuracy as the last step is not
   carried out.

The reservoir tip does not need to be
filled completely.

      The 0.5 ml, l, 1.25 ml and 2.5 ml tips

      may be used with a 100 

µ

l pipette tip

      affixed to the end. This facilitates filling
      from vessels with narrow necks and
      provides better accuracy with small
      volumes.

5. Dispensing

 Check that the volume setting is correct.

 Wipe the end of the tip with a lint free wipe or other
 appropriate material.

 Place the tip at an angle to the wall of the vessel.

    Press the plunger down to the stop to dispense.

After using the 25 ml and 50 ml
tips, the adapter can be removed
by turning it counter-clockwise and
reused with another tip.

Avoid touching the end of the tip
when removing it from the pipette.

       When using viscous solutions,
      the fit of the tip may loosen after
      several pipettings.