LABNET Excel 

Single and Multi-channel Electronic Pipettors 

 
 

Operation Manual 

Version: 2.9 

 

 

 

 

 

 

 

www.Labnetlink.com  

 

 

 

 

 

Содержание Excel E10-1

Страница 1: ...LABNET Excel Single and Multi channel Electronic Pipettors Operation Manual Version 2 9 www Labnetlink com...

Страница 2: ...abnet Excel Single Multi channel Electronic Pipettors All models comply with the following European standards EMC EN 55011 Class A EN 60601 1 2 To the best of my knowledge and belief these units confo...

Страница 3: ...ting the Battery 5 3 3 Charging the Battery 6 3 4 Start Pipetting 8 3 5 Manual Blow out Operation 10 3 6 Hanger Installation 10 4 LABNET Excel electronic pipettor Overview 11 4 1 Keypad Function 12 4...

Страница 4: ...Aspirating Mode 23 10 2 SD Sequential Dispensing Mode 24 11 Sterilization 26 11 1 Removal of the Lower Piston Assembly 27 11 2 Reattach the lower piston assembly 28 12 Maintenance 30 13 Troubleshooti...

Страница 5: ...nnel only Contact LABNET International Inc for information on authorized service options 4 Use original spare parts and accessories only 1 2 Battery Safety The LABNET Excel electronic pipettors operat...

Страница 6: ...the specified battery charging unit or DC in jack to charge the battery 9 Do not use a leaking battery 10 If charging is not completed within the specified time period unplug the charger and discontin...

Страница 7: ...e on a single battery charge z Ergonomic design z Automatic calibration z User friendly graphic operation interface z Adjustable tip ejector z Automatically remembers the last used pipetting protocol...

Страница 8: ...0 1 0 2 0 0 3 E200 8 8 10 200 1 2 0 0 6 1 0 0 15 E1200 8 8 100 1 200 1 1 5 0 5 0 6 0 15 Note LABNET Excel electronic pipette Accuracy and Precision Specifications have been attained using LABNET Exce...

Страница 9: ...ording to your order please contact your distributor or sales representatives for replacement immediately Please charge the battery before first time operation Remove the battery cover by pressing the...

Страница 10: ...e power plug of the power adaptor with the external power source Connecting the power plug before inserting the battery will inactivate the charging process The indication bar inside battery symbol wi...

Страница 11: ...o charge the battery Figure 4 Using the optional charging kit to charge the battery Note 1 For safety reasons the LABNET Excel electronic pipettors cannot be used during charging 2 Please do not over...

Страница 12: ...Excel electronic pipettor will be automatically powered off if it s not in use for more than 10 minutes Select the desired function mode and settings refer to the following chapters 6 10 before pipet...

Страница 13: ...nteed only when using LABNET Excel tips Contact LABNET International for pipette tip ordering information The LABNET Excel electronic pipettor is an ergonomic instrument designed to reduce hand fatigu...

Страница 14: ...uid The piston will then pause and return to its starting position Alernatively the user can press and hold the PLUNGER button to dispense The piston will travel the extra distance to blow out remaini...

Страница 15: ...SET Button Tip Ejector Keypad LCD Display PLUNGER Button DC In Battery Cover Figure 7 1 channel LABNET Excel electronic pipettor overview This section presents an overview of the LABNET Excel electron...

Страница 16: ...ble Singals Count No Prog No Figure 8 LCD display Panel 4 1 Keypad Function Button Symbol Function Description Set Enter Confirms your choice Exit Esc Exits to previous mode setting Up Increases volum...

Страница 17: ...re can speed up the setting processes 4 2 Audible Signals Audible Sound Operation Low tone beep Aspirating High tone beep Dispensing Tip Ejection Error codes 2 beeps MD MIX SE mode cycle or RESET rout...

Страница 18: ...re 9 Function Map SYS AUTO MD MIX SE Function Mode Aspirating Speed Dispensing Number COUNT Basic Aspirating Dispensing Mix Number COUNT Dispensing Speed Audible Signals Program Recall Reverse Pipetti...

Страница 19: ...to confirm the selection of mode and the aspirating speed symbol will start flashing 3 Press Up Down to change the aspirating speed Slowest Fastest 4 Press SET to confirm the speed selection and move...

Страница 20: ...ce On Off 3 Press SET to confirm the selection and move to the Tip Ejection Alarm setting 6 4 Blow out 1 The Blow out symbol should now be flashing 2 Press Up Down to select the preference On Off Plea...

Страница 21: ...and remain there until the PLUNGER button is released 6 5 Reverse Pipetting 1 The Reverse Pipetting symbol should now be flashing 2 Press Up Down to choose the preference On Off 3 Press SET to confirm...

Страница 22: ...flashing 2 Press SET repeatedly to confirm the selection of all 6 function settings until the symbol starts flashing 3 Press Up Down to select the PROG number 4 Press SET to confirm the selection 5 P...

Страница 23: ...r the mode selection which causes the symbol to start flashing If not press Up Down to select AUTO mode 2 Press SET to confirm the mode selection 3 Use Up Down to select the desired pipetting volume 4...

Страница 24: ...ycle counts 6 Press SET to confirm selected mixing cycle counts The mixing volume will start flashing 7 Press Up Down to select the desired mixing volume 8 Press SET to confirm selected mixing volume...

Страница 25: ...to dispense the samples according to the pipetting counts For the last count you will hear two beeps to indicate that the next dispense is the last and the display will show to remind the user to pres...

Страница 26: ...etting volume will start flashing 6 Use Up Down to select the desired mixing volume 7 Press SET to confirm the selection The total mixing volume and counts will be displayed Press the PLUNGER button t...

Страница 27: ...lumes The total aspirating volume including air gap volume cannot exceed maximum volume range 9 Press SET to confirm the selection 10 The count 01 and aspirating volume will appear on the display The...

Страница 28: ...aspiration mode 2 1200ul models do not include the Sequential Aspirating SA function because the air gap is not strong enough to support the weight of samples 3 Sequential Aspirating SA function will...

Страница 29: ...spiration is completed the dispensing count and volume will appear and will flash to indicate that the LABNET Excel electronic pipettor is ready for dispensing 11 The LABNET Excel electronic pipettor...

Страница 30: ...amage the manifold and void the warranty The autoclaved parts must be allowed to dry completely at room temperature for at least 2 hours before reassembly Users can follow Figures 11 14 to disassemble...

Страница 31: ...n as shown in Figure 11 3 Pull the lower piston assembly downward slowly until the metal cylinder lock appears as shown in Figure 12 4 Lay the LABNET Excel electronic pipettor down on a flat surface P...

Страница 32: ...eattach the lower piston assembly of LABNET Excel electronic pipettor 1 Press the PLUNGER button or any button of the keypad to make the step motor shaft retract about 1cm back into the handle The ste...

Страница 33: ...29 3 Press the RESET button The step motor shaft will connect with the piston of the lower part automatically After the connection is completed the LABNET Excel electronic pipettor will beep twice...

Страница 34: ...e the pipette more frequently Lubrication of the piston should be performed only by authorized LABNET service personnel or trained users and only the proper LABNET Excel electronic pipettor lubricant...

Страница 35: ...niform wetting of the plastic tip Rinse the tip once prior to pipetting To ensure the product s quality and performance the LABNET Excel electronic pipettor is designed with a self diagnosis program T...

Страница 36: ...material blocking the hole at bottom of the cone Use MIX mode and distilled water to wash Air dry Failure to aspirate Piston movement is blocked Lubricate piston Bad battery contact Reinsert the batte...

Страница 37: ...e 5 50o C Operating Humidity RH 0 85 Battery 900 mAh 3 6 V or above Li Ion Battery Charger Kit Optional Carrousel Stand Optional Certification Complies with CE Class A EN60101 1 2 EN50082 1 EN 55011 C...

Страница 38: ...oes not cover defects caused by excessive wear and tear or damage due to shipping accident abuse misuse problems with electrical power or usage not in accordance with product instructions or if other...

Страница 39: ...35 Note...

Страница 40: ...36 Note...