Labnet Enduro™ Power Supplies

Instruction Manual

This manual is available in additional languages at 

www.labnetlink.com

.

Catalog Numbers: 

High Amperage (250V)

E0203
E0203-230V

General Purpose (300V)

E0303
E0303-230V

Содержание Enduro E0203

Страница 1: ...et Enduro Power Supplies Instruction Manual This manual is available in additional languages at www labnetlink com Catalog Numbers High Amperage 250V E0203 E0203 230V General Purpose 300V E0303 E0303...

Страница 2: ...ion of Buttons and Switches 3 7 0 Getting Started 4 8 0 Recommendations 4 9 0 Operating Instructions 5 9 1 Constant Operation Protocol 5 9 2 Basic Programming Protocol 250V only 6 10 0 Choosing Limiti...

Страница 3: ...nt and power range of the Power Supply To prevent accidental electric shocks to the user always use gel box systems that have safety lids Labnet recommends against the use of gel box systems and or po...

Страница 4: ...300W 90W Output voltage range 5 to 250V 2 to 300V Output current range 10 to 3000 mA 4 to 500 mA Operating modes Constant voltage Constant current Constant power 1V step 10 mA step 1W step 1V step 1...

Страница 5: ...t Enduro 300V Power Supply is designed to efficiently handle multiple electrophoresis gel tanks and use a small amount of lab space This instrument is ideal for DNA RNA electrophoresis and mini PAGE g...

Страница 6: ...the power switch and the AC inlet located on the rear of the unit are easily accessible 4 Ensure the AC power switch is in the Off position 5 Attach the power cord to the AC inlet Use only properly gr...

Страница 7: ...e first line on the top right and the non constant value is displayed in the second line on the right side of the display screen 2 Press Constant button to choose constant volt or amperage Choosing co...

Страница 8: ...ided below We recommend reading the guidelines provided in this manual for best results before starting an operation The Programmable Mode allows you to vary levels in voltage V current mA and power W...

Страница 9: ...ethods resistance increases throughout the run Limiting the voltage provides the following advantages Current and power decrease throughout the run providing an improving margin of safety over time Th...

Страница 10: ...t the condition by making sure the tape is removed from the pre cast gel or buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit High voltage applic...

Страница 11: ...ent P I x V 13 0 Common Errors Found with Electrophoresis Power Supplies No load The electrophoresis system is not connected to the power leads check the power leads The electrophoresis system has a s...

Страница 12: ...orning of any such defect Corning is not liable for any incidental or consequential damages commercial loss or any other damages from the use of this product This warranty is valid only if the product...

Страница 13: ...___________________________________ 16 0 Equipment Disposal According to Directive 2012 19 EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment...

Страница 14: ...mation visit www corning com lifesciences or call 800 492 1110 Outside the United States call 1 978 442 2200 or contact your local Corning sales office Corning Incorporated Life Sciences 836 North St...