background image

This guide provides site preparation information for the Echo® 650 Series Liquid Handler. This includes models 650, 655, 650T
and 655T. The topics below describe the site preparation requirements and instrument specifications for the Echo Liquid
Handlers.

Pre-Installation Requirements

l

Site Requirements

l

User Supplied Equipment Requirements

l

Communication Requirements

l

Accommodating Your Echo Liquid Handler

l

Noise Ratings

l

Power Ratings

What You Will Be Receiving

l

Shipping Dimensions

l

Echo Liquid Handler Components

l

Chiller

l

Junair Compressor (Optional and sold separately by Labcyte)

l

Vacuum Pump (Optional and sold separately by Labcyte)

l

Client PC (Sold separately by Labcyte)

Pre-Installation Requirements Verification Form

LABCYTE INC.

SITE PREP GUIDE

|

Echo® 650 Series Liquid Handler

© 2019 LABCYTE INC. All rights reserved.

QSG-2019-1

FEB 2019

Echo® 650 Series Liquid Handler
(Models: 650, 655, 650T, 655T)

SITE PREP GUIDE

Server Version 3.1 | FEBRUARY 2019

Содержание Echo 650 Series

Страница 1: ...modating Your Echo Liquid Handler l Noise Ratings l Power Ratings What You Will Be Receiving l Shipping Dimensions l Echo Liquid Handler Components l Chiller l Junair Compressor Optional and sold separately by Labcyte l Vacuum Pump Optional and sold separately by Labcyte l Client PC Sold separately by Labcyte Pre Installation Requirements Verification Form LABCYTE INC SITE PREP GUIDE Echo 650 Seri...

Страница 2: ...eet of the Echo Liquid Handler supporting a minimum of 800 W are required The Echo Liquid Handler and SMC chiller maximum power consumption is 600W and 500W respectively Up to four additional electrical sockets are required for the client PC monitor vacuum pump optional and air compressor optional Note The Echo Liquid Handler power configuration is factory set and requires a Field Service visit in...

Страница 3: ...sters If the table has casters they should be locked If the chiller will be installed on the bench surface it must accommodate an additional 14 kg 31 lbs Destination plate samples At installation ensure common destination plates to be used with the instrument are available for tuning the destination plate gripper Coupling Fluid Use only ASTM D 1193 Type 1 water in the fluidics system Coupling Flui...

Страница 4: ... onto a workstation it must contain an Intel Core i5 processor or better Note If you are setting up an Echo 650T or 655T Liquid Handler that will be using tubes a capper decapper is required before Labcyte Service and Support can install or service the instrument 4 SITE PREP GUIDE Echo 650 Series Liquid Handler PN 001 16480 ...

Страница 5: ...by robot Side It is recommended to have 121 9 cm 48 in minimum on left and right sides to allow for serviceability The Echo Liquid Handler should have sufficient space to be rotated otherwise it will need to be removable for service Rear It is recommended to have 30 5 cm 12 in minimum behind the instrument to allow for cables hoses tubing and maintenance access The coupling fluid bottle and option...

Страница 6: ...oise ratings for the Echo Liquid Handler are below Noise Rating 64 dBA Power Ratings The power ratings for the Echo Liquid Handler are below Power Rating 600W without the chiller Heat Rating 600W without the chiller 6 SITE PREP GUIDE Echo 650 Series Liquid Handler PN 001 16480 ...

Страница 7: ...ir line 4 6 meters 15 ft of polyurethane tubing 6 35 mm 0 25 in OD 4 24 mm 0 170 in ID SMC Pneumatics P N TIUB07 BU LCD 13004 coupling body Colder Products Coupling fluid bottle 1 L P N 001 5438 Accessory kit Includes tubing and power cord for connecting the chiller P N 001 15272 SMC chiller P N 001 11149 Ethernet cable Crossover 3 meters 10 ft Software Echo Client Utility installer provided on CD...

Страница 8: ...uses an auto switching power supply capable of running any single phase AC voltage in the range 100 240V 50 60 Hz A power cord suitable for the install location is provided Maximum power is 500W Temperature Control The SMC chiller controls the temperature to 0 02 C by cooling and heating as appropriate The SMC chiller is air cooled Ambient Operating Conditions Temperature 10 35 C Relative Humidity...

Страница 9: ... 60 60 50 60 50 60 50 60 Motor HP 0 60 0 60 0 60 0 60 0 60 0 60 0 60 0 60 0 60 kW 0 44 0 44 0 44 0 44 0 44 0 44 0 44 0 44 0 44 Displacement I min 108 138 138 108 138 108 138 108 138 CFM 3 81 4 87 4 87 3 81 4 87 3 81 4 87 3 81 4 87 FAD 8 bar l min 38 44 44 38 44 38 44 38 44 CFM 1 34 1 55 1 55 1 34 1 55 1 34 1 55 1 34 1 55 Maximum pressure bar 8 8 8 8 8 8 8 8 8 psi 120 120 120 120 120 120 120 120 12...

Страница 10: ...Figure 3 Junair OF302 4B oil less compressor 10 SITE PREP GUIDE Echo 650 Series Liquid Handler PN 001 16480 ...

Страница 11: ... the specifications for the Vacuubrand MZ 2S NT vacuum system Permissible ambient temperature storage operation 14 to 140 50 104 F 10 to 60 10 40 C Permissible relative atmospheric moisture during operation no condensation 30 to 85 No load speed 50 60 hZ 1500 1800 rpm Device fuse Slow blow fuse 6 3 A Maximum pumping speed 50 60 Hz 1 3 1 4 cfm 2 2 2 4 m3 h Ultimate vacuum absolute 5 2 Torr 7 mbar M...

Страница 12: ...Two Ethernet Cards Included Accessories l Microsoft Office 2016 l Monitor 22 in DVI D DP VGA l Dell USB Keyboard Black l Dell MS111 USB Optical Mouse OptiPlex and Fixed Precision l Labcyte Mousepad Software specifications Optionally the following Echo software applications are available for purchase l Echo Plate Reformat l Echo Dose Response l Echo Cherry Pick l Echo Array Maker l Echo Plate Audit...

Страница 13: ...ab environment setup or required equipment please contact your local Labcyte Representative or email support labcyte com LABCYTE INC SITE PREP GUIDE Echo 650 Series Liquid Handler FOR RESEARCH USE ONLY Not for use in diagnostic procedures QSG 2019 1 FEB 2019 LABCYTE INC 170 Rose Orchard Way San Jose CA 95134 USA Toll free 1 877 742 6548 Fax 1 408 747 2010 SUPPORT North America Europe Japan Asia Ot...

Отзывы: