System Description
3-19
3.3.3 Environmental
Temperature, operating:
21°C ± 5°C (70°F± 9°F)
Temperature, storage:
5°C
–
45°C (41°F
–
113°F)
Humidity, operating: 10
–
80%, non-condensing
Mounting surface:
Sturdy, stable, capable of handling 128 kg (283 lbs)
weight
Pollution degree:
2
Installation category:
II
3.3.4 Electrical
There are two electrical configurations of the Echo liquid handler:
3.3.5 Client PC
Operating system:
Microsoft
®
Windows XP SP3 or Windows 7
using 32-or 64-bit OS; Windows Net Framework 2.0
CPU:
Intel
®
Core™ 2 Duo or later
Memory:
1GB or greater
Hard drive storage:
1GB
Video:
1280 x 1024 resolution
API:
ActiveX
Programming languages:
C, C++, C#, Visual Basic, Visual J#, or
Windows Net Framework compatible
3.3.6 Communications
Network connection:
10/100/1000 BaseT
Network protocol:
TCP/IP
User account:
Labcyte (The client PC must have a Labcyte account, with
administrator privileges, that is accessible by LabcyteService and Support to
service the Echo liquid handler system.)
Config.
Voltage
Frequency
Power
Fuse
A
100–120 VAC
50/60 Hz
10 A
250 VAC, 10 A, Fast Blow
B
200–240 VAC
50/60 Hz
5 A
250 VAC, 5 A, Fast Blow
Содержание Echo 520
Страница 1: ...For Echo models 520 550 555 ...
Страница 20: ...xx Echo Liquid Handler User Manual ...
Страница 58: ...2 14 Echo Liquid Handler User Manual ...
Страница 82: ...3 24 Echo Liquid Handler User Manual ...
Страница 100: ...4 18 Echo Liquid Handler User Manual ...
Страница 160: ...5 60 Echo Liquid Handler User Manual ...
Страница 192: ...A 8 Echo Liquid Handler User Manual ...
Страница 199: ......