Installation and Relocation
2-5
2.2
Installation overview
The Labcyte Echo liquid handler is installed by Labcyte field engineers. The
following is a brief overview of the hardware and software installation
steps they perform:
1.
Unpack the Echo liquid handler and place it in the area that meets the
minimum operating space requirement.
2.
Connect the Echo liquid handler to AC power, air pressure, vacuum
pressure, surge tank, coupling fluid bottle, and the coupling fluid chiller.
3.
Fill the coupling fluid bottle and prime the coupling fluid circuit.
4.
Perform a basic functional check of the Echo liquid handler.
5.
Perform the Site Acceptance Test (SAT) process (optional).
6.
Change the Echo IP address, if a different IP address is required by your
system administrator.
7.
Install the Echo software onto the customer-supplied PC
(client workstation).
8.
Connect the client PC to the Echo liquid handler and verify
communications.
Figure 2.2 Installed system
Ethernet cable
Compressed air line
Vacuum line
Client PC
Chiller
Echo liquid handler
fluid lines
Chiller
Bench
Floor
(cross-over)
Surge tank