Figure 11: Workcell Tab - Yellow Icon Indicates Open Door
Interlock Sensors
The Access DRS has 16 interlock sensors for enclosure access panels including doors, blanking panels, tunnel doors.The system
relies on the sensors to guarantee a safe operation of the system at all times. The impact of a triggered sensor on the system
during operation or when it is idle depends on the location and the type of sensor.
All sensor events are logged and communicated through Tempo software (Alarms Tab) and the integrated light tower system
alerts.
This table lists operator actions that trigger a sensor and how that impacts the Access system.
Actions That Trigger
Sensors
Impact on the Access System
Device door is
opened
l
Robots are slowed into safety mode
Device shelf is pulled
l
Robots are slowed into safety mode
l
Devices on the shelf are switched to Standalone mode
Table 4: Interlock Sensor Action And Result
Interlock Sensors
CHAPTER 3 | Safety Warnings
19
Содержание Access Dual Robot
Страница 25: ...Figure 13 Device Module 1 Right Top View Device Module 1 CHAPTER 5 System Dimensions 25 ...
Страница 27: ...Figure 15 Device Module 2 Top View Device Module 2 CHAPTER 5 System Dimensions 27 ...
Страница 29: ...Figure 17 Echo Module Top View Echo Module CHAPTER 5 System Dimensions 29 ...
Страница 31: ...Figure 19 Storage Module AmbiStore Top View AmbiStore Storage Module CHAPTER 5 System Dimensions 31 ...
Страница 33: ...Figure 21 Storage Module MicroServe Top View MicroServe Storage Module CHAPTER 5 System Dimensions 33 ...
Страница 35: ...Figure 23 Storage Module Cytomat Top View Cytomat Storage Module CHAPTER 5 System Dimensions 35 ...
Страница 71: ...Figure 70 Sample Error Screen Robot Device Recovering From An Emergency Shutdown CHAPTER 7 Basic Operations 71 ...