background image

 

 

 
 

 

 
 

 

 

 
 
 

 

 
 
 
 
 

 
 
 
 

 

 

 

 

 

 

 

 
 

 

 
 
 

LABCONCO CORPORATION 

 

8811 Prospect Avenue 
Kansas City, MO 64132 
 
(800) 821-5525  I  +1 (816) 333-8811 
 
labconco@

labconco.com 

 

 

Please read this manual before 

operating equipment 

 

Original Instructions

 

User’s Manual 

Содержание Purifier Logic+ 30348

Страница 1: ...ABCONCO CORPORATION 8811 Prospect Avenue Kansas City MO 64132 800 821 5525 I 1 816 333 8811 labconco labconco com Please read this manual before operating equipment Original Instructions User s Manual...

Страница 2: ...2 Logic Type B2 Biosafety Cabinets 2019 Present Catalog Numbers Ending in x1 30348xxxx 30368xxxx...

Страница 3: ...30 p m Central Standard Time Part 3848331 Rev C ECO L847 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a resul...

Страница 4: ...ng the Cabinet on a Labconco Base Stand 16 Telescoping Base Stands 16 Manual or Electric Hydraulic Lift Base Stands 16 Exhaust System Connections 17 18 Exhaust System Requirements 18 Connecting the Ca...

Страница 5: ...the Security Lock 41 Setting the USB Output Rate 41 Navigating UV Parameters 42 The Tools Submenu 42 Timer Operation 43 Interval Timer Operation 43 Stopwatch Timer Operation 43 If An Airflow Alert Act...

Страница 6: ...nditions 62 APPENDIX D ACCESSORIES 63 APPENDIX E QUICK CHART 64 CAUTION See Manual When this symbol is on the unit it indicates a caution that is detailed in this manual ATTENTION Voir manuel Lorsque...

Страница 7: ...s manual and keep it handy for future reference If you are unfamiliar with how biosafety cabinets operate please review Section 4 Performance Features and Safety Precautions before you begin working i...

Страница 8: ...power must be located near the installation site This section lists all requirements for Clearance Location Exhaust Electrical Service Lines Refer to Appendix C Specifications for complete cabinet ele...

Страница 9: ...n closed Figure 2 1 shows the preferred location for the cabinet Preferred location Air register blocked or redirected to prevent cabinet disruption Alternate location Main Disconnect Device Power cor...

Страница 10: ...he cabinet s exhaust port should be clear of structural elements water and utility lines or other fixed obstructions There should be enough clearance to allow for the passage of a 10 stainless steel d...

Страница 11: ...rical codes for properly rated circuit breakers For safe operation the dedicated outlet must provide the protective earthing ground connection to the cabinet On 100 and 115 VAC models both internal el...

Страница 12: ...mable gases or solvents should be avoided in the cabinet Open flame in the cabinet will disrupt the laminar airflow in the cabinet and may damage the HEPA filters Flammable gases or solvents may reach...

Страница 13: ...e lines Connect to an exhaust system optional Arrange certification of the cabinet Tools required for installation of the cabinet include two 1 2 wrenches a flat blade screwdriver a 2 Phillips screwdr...

Страница 14: ...components have been checked installed and tested 2 The cabinet is secured to the pallet in two places on each side To access the nuts and bolts remove the side panels by removing and keeping the two...

Страница 15: ...tached to its pallet by using a floor jack or a furniture dolly underneath the unit DO NOT move the cabinet by tilting it onto a hand truck When lifting the cabinet DO NOT lift the cabinet in the midd...

Страница 16: ...tric pump that drives the hydraulic legs of the stands All of the hydraulic stands are equipped with fixed feet but can be converted to caster wheels with the Caster Kit catalog number 3784000 The bas...

Страница 17: ...rk is appropriate for the cabinet Type B2 Exhaust system must be reliable maintained frequently inspected and preferably redundant Exhaust termination must be distant from other building air intake sy...

Страница 18: ...RS AND A BACKDRAFT DAMPER WHICH ARE SUITABLE FOR MOST SINGLE CABINET INSTALLATIONS SEE APPENDIX D Connecting the Cabinet to Utility Service Lines Note Some models have a solenoid valve connected to th...

Страница 19: ...the fitting 5 Tighten the tube fitting nut hand tight and then using a 7 16 inch wrench tighten it at least turn more 6 Close the service valve in the cabinet and then slowly open the shutoff valve on...

Страница 20: ...a spatula or similar device between the sealing plug and the body of the portal and prying the plug out 2 Locate the pass thru universal sealing plug which is included in the parts box located underne...

Страница 21: ...et 3 Using a putty knife remove and discard the stainless steel cover that is sealed over the drain mounting holes Scrape off remaining sealant that is around the holes 4 Apply a light coating of sili...

Страница 22: ...umber 49 in effect when the cabinet was manufactured Downflow Velocity Profile Test Inflow Velocity Test Airflow Smoke Patterns HEPA Filter Leak Test Optional Canopy Alarm Test and Operation Vibration...

Страница 23: ...material or media is typically made of borosilicate microfibers formed into a thin sheet in a process similar to the production of paper This sheet is folded or pleated to increase its surface area T...

Страница 24: ...A Filters are only effective against particulate material Gases will pass through the filter Filter frame Polyurethane seal Between media pack and filter frame Filter media pack A continuous sheet of...

Страница 25: ...tice the laminar downflow of air in the cabinet captures any aerosol generated in the work area of the cabinet and directs it to the HEPA filters In order to be true laminar downflow a number of indiv...

Страница 26: ...is drawn into the front of the cabinet at the front grille This curtain of air makes it more difficult for aerosols to escape out of the work area of the cabinet and into the outside environment This...

Страница 27: ...ming alternating current to direct current as well as its own microprocessor to control and measure the motor s operation The motor utilizes Labconco s exclusive Constant Airflow ProfileTM CAP program...

Страница 28: ...er to be effective the UV light must directly strike the nucleic acid and its effectiveness can be diminished or negated by dissolved proteins or metals or by other UV opaque substances protecting the...

Страница 29: ...nce with local and national electrical codes Failure to do so may create a fire or electrical hazard Do not remove or service any electrical components without first disconnecting the cabinet from ele...

Страница 30: ...ntilation Carefully follow the disinfectant s safety instructions Always dispose of disinfecting solutions in accordance with local and national laws DO NOT allow disinfectants with high concentration...

Страница 31: ...set switch for resetting its microprocessors The switch is located on the front of the electronics module on top of the cabinet as shown in Figure 5 1 Ensure that the switch is in the ON up position b...

Страница 32: ...t status and inflow velocity as shown in Figure 5 3 In the event of an alarm the Information Center will immediately display a context sensitive message indicating the cause of the alarm and possible...

Страница 33: ...urn sash to proper working height Airflow Alert The airflow patterns in the cabinet have changed resulting in a sudden change in the motor speed This is most likely due to a blockage of the grille or...

Страница 34: ...flow after 60 seconds the exhaust alarm will be activated Exhaust Alarm If the facility s exhaust system does not achieve proper exhaust airflow volume within the 60 second start up window or the faci...

Страница 35: ...ystem senses when the sash is closed and allows the optional ultraviolet UV lamp to operate only when the sash is closed to protect the operator from irradiation Starting the Cabinet 1 To start the ca...

Страница 36: ...at any time overrides the automatic operation LIGHT Turns the LED lamps on or off Closing the sash automatically turns the lights off When the lights are in automatic SmartStart mode raising the sash...

Страница 37: ...l override Automatic Operation mode functions such as SmartStart and NightSmart To access the main menu press the MENU button The display panel will change to the main menu To select from the various...

Страница 38: ...bled an audible tone will sound during cabinet power up This also enables or disables audible tones from the keypad any button press Selecting a Language UP and DOWN will move among the selectable lan...

Страница 39: ...M PM format or 24 Hour format The selected field Hours or Minutes will flash set the current time using UP and DOWN Hours will flash first once correct use OK MUTE to switch to Minutes Note AM or PM w...

Страница 40: ...anually turned on from the keypad When OK Mute is pressed the next configuration screen will appear if your cabinet is configured for a UV lamp If your Cabinet is configured for a UV light you will se...

Страница 41: ...immediately after the blower is turned off The security lock is deactivated by holding DOWN for three seconds If blower is not turned on within 5 minutes of unlocking the keypad will relock The featur...

Страница 42: ...or 3 seconds To change the desired UV lamp lifetime select Lifetime as seen in Fig 5 7p then OK MUTE The screen shown in Fig 5 7q will display To change the UV lamp lifetime number of operating hours...

Страница 43: ...imer interval 3 When the proper interval is selected press OK Mute to start the timer 4 When the timer reaches 00 00 an audible alarm will sound and the timer will reset itself and repeat the countdow...

Страница 44: ...Centers for Disease Control and Prevention cdc gov biosafety publications Planning Thoroughly understand procedures and equipment required before beginning work Arrange for minimal disruptions such a...

Страница 45: ...Techniques Keep all materials at least 4 inches 100 mm inside from the sash and perform all contaminated operations as far to the rear of the work area as possible Segregate all clean and contaminated...

Страница 46: ...d be covered before being removed from the cabinet Wipe Down Wipe down the interior surfaces of the cabinet with a suitable disinfectant or 70 ethanol and allowed to dry Periodically lift the work sur...

Страница 47: ...tions and specifications are published in a separate publication Technical Manual Purifier Logic Cabinets Second Generation This manual is available from Labconco s website labconco com A complete cer...

Страница 48: ...r Operate cabinet blower noting Filter Life percentage in log Using 70 alcohol solution wipe down cabinet exterior Disinfect and lift work surface Surface disinfect the area beneath with suitable chem...

Страница 49: ...ure to engage the tabs on the back corners of the work surface with the slots on the rear wall of the work area Front Grille Removal Note The grille must be thoroughly decontaminated before removing i...

Страница 50: ...you will feel some resistance Note the orientation of the towel catch 3 Surface decontaminate the towel catch before removing it 4 Reinstall the towel catch by sliding it back into position in the cor...

Страница 51: ...ated on the bottom corners of the dress panel 2 Swing the bottom of the dress panel out to clear the LED lamps and then lift the dress panel straight up and away from the cabinet 3 To reinstall the pa...

Страница 52: ...e 6 3 3 Locate the Left End Cap that aligns both LED lamps Fig 6 5a remove the Left End Cap by pulling it away from the lamp ends 4 Remove both Socket Caps on the right end of each LED lamp by pulling...

Страница 53: ...align rotationally with the Left End Cap This ensures the dead band stripe is positioned correctly for maximum cabinet lighting Changing the Optional UV Lamp Note For optimum performance the UV lamp s...

Страница 54: ...setting a Circuit Breaker To reset any of the circuit breakers located on the left side of the electronics module depress the white button until it sets Circuit Breakers The front breaker protects the...

Страница 55: ...e cabinet into appropriate electrical service Check connection to control box on top of cabinet System Reset Switch is Off Turn on the System Reset Switch Circuit breaker s tripped Reset circuit break...

Страница 56: ...s are defective Replace defective lamp s Lamp wiring is disconnected Inspect lamp wiring UV light not working Sash is open Close sash UV light will not work with the sash open Lamp is defective Repla...

Страница 57: ...he work surface are unobstructed Exhaust Airflow Check goes off and alarm sounds 30 seconds after the blower is started Roof blower is off Ensure that roof blower is working Exhaust damper is improper...

Страница 58: ...r slots or grille blockage of the exhaust outlet Ensure that all return air slots grilles and the exhaust outlet are unobstructed External factors are disrupting the cabinet airflow patterns or acting...

Страница 59: ...iption 1 1 3850500 Pre Filter for 4 Foot B2 Model 1 3850501 Pre Filter for 6 Foot B2 Model 2 1 3438501 Exhaust HEPA Filter 4 Foot B2 Model 1 3438503 Exhaust HEPA Filter 6 Foot B2 Model 3 1 3838401 Sup...

Страница 60: ...are shown in inches B 1 Width A B C 4 54 3 48 5 17 2 6 78 3 72 5 28 9 A B 72 6 21 7 Max Loading C 8 8 Exhaust duct stub sized for 10 inch outside diameter duct connection Power Cord Connection Power C...

Страница 61: ...036xxx01 5 A 115 V 60 Hz 12 A 3036xxx21 5 A 100 V 50 60 Hz 12 A 3036xxx 11 31 41 51 61 71 2 5 A 230 V 50 60 Hz 6 A Motor Specifications Cabinet Electrical Requirements All Cabinets all Voltages 1 2 H...

Страница 62: ...ceed 10 of the nominal voltage Transient overvoltages according to Installation Categories II Overvoltage Categories per IEC 1010 Temporary voltage spikes on the AC input line that may be as high as 1...

Страница 63: ...63 Appendix D Accessories Labconco offers a full line of accessories to enhance your cabinet s operation and usability For a complete list visit labconco com...

Страница 64: ...The primary serial tag is on the lower right edge of the front dress panel The secondary serial tag is located on the front of the electronics module on the top right side of the cabinet The first two...

Отзывы: