95
QuickChart
– Type C1
Type C1 Cabinet Width
4-ft
6-ft
Sash Opening (inches)
8
10
8
10
Starting Serial #
2006_
2006_
2006_
2006_
Downflow Data
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
Grid points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.75
5.55
5.75
5.55
Distance between columns
5.21
5.21
5.50
5.50
Inflow/Exhaust Data
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
2.69
3.37
4.03
5.03
Nominal Avg. Inflow Vol. (CFM)
283
354
423
529
Avg. Inflow Vol. Range (CFM)
269-296
337-371
403-443
503-553
Secondary Inflow Data
Sash Opening Template
1
4+
4+
6+
6+
Sensor distance (inches)
2
3.25
3.25
4
4
# of Test points
8
8
12
12
Avg. Inflow Vel. Range (FPM)
249-274
312-344
202-222
252-277
Correction Factor (CF)
1.08
1.08
2
2
Corrected Inflow Volume (CIV)
(Avg. velocity x CF)
269-296
337-371
403-443
503-553
Supply HEPA Filter Leak Test Data
Air Displacement (CFM)
520
520
665
665
Laskin Nozzles needed
1
1
1
1
Theoretical aerosol conc. (ug/l)
3
26
26
20
20
Actual aerosol conc. (ug/l)
4
13
13
12
12
Supply HEPA Data
Labconco P/N
3838401
3838401
3838403
3838403
Width x Depth x Height (in.)
5
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Pressure Drop (in. H
2
O)
.5+/-
.1”
.5+/-
.1”
.38-
.5”
.38-
.5”
Exhaust HEPA Filter Leak Test Data
Air Displacement (CFM)
283
354
423
529
Laskin Nozzles needed
1
1
1
1
Theoretical aerosol conc. (ug/l)
3
48
38
32
26
Actual aerosol conc. (ug/l)
4
28
22
19
15
Exhaust HEPA Data
Labconco P/N
3838501
3838501
3838503
3838503
Width x Depth x Height (in.)
5
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
353
353
529
529
Performance (in. H
2
O)
.2-.35
”
.2-.35
”
.2-.35
”
.2-.35
”
LED Lamp Data (2 each)
7
Labconco P/N
1297504
1297504
1297506
1297506
LED Lamps (Direct Drive)
15T8-48
15T8-48
32T8-72
32T8-72
Color (
o
K)
4000
4000
4000
4000
Lumens
1850
1850
3200
3200
Glass Type
Frosted
Frosted
Frosted
Frosted
Remote Exhaust System Data
12
DIM Total Exhaust Volume
C Canopy (CFM)
8
323
404
473
579
Traverse Total Exhaust Volume Cabinet
+ Canopy (CFM)
9
366
444
520
637
IRV
11
Slot Area (ft
2
)
.12
.12
.12
.12
Nominal IRV Inflow (CFM)
50
50
50
50
Nominal IRV Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Recommended Duct Vacuum (in. H
2
O)
10
0.3
0.3
0.3
0.3
Volume of Hull
Cubic Feet
39
39
58
58
See footnotes on following page
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...