![Labconco Logic+ A2 Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/labconco/logic-a2/logic-a2_technical-manual_3395427090.webp)
90
QuickChart
– Type A2 3-ft & 4-ft
Type A2 Cabinet Width
3-ft
4-ft
Sash Opening (inches)
8
9
10
8
9
10
12
Starting Serial #
1905_
1905_
1905_
1905_
1905_
1905_
2006_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
55+/-5
55+/-5
60+/-5
Grid points (rows x columns)
21 (3x7)
21 (3x7)
21 (3x7)
24 (3x8)
24 (3x8)
24 (3x8)
24 (3x8)
Grid distance from back & sides
6.0
6.0
6.0
6.0
6.0
6.0
6.0
Distance between rows
5.75
5.75
5.55
5.75
5.75
5.55
5.44
Distance between columns
4.08
4.08
4.08
5.21
5.21
5.21
5.21
Inflow/Exhaust Data
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
2.03
2.28
2.53
2.69
3.03
3.37
4.04
Nominal Avg. Inflow Vol. (CFM)
213
240
266
283
318
354
424
Avg. Inflow Vol. Range (CFM)
203-223
228-251
253-278
269-296
303-333
337-371
404-444
Secondary Inflow Data
Sash Opening Template
1
3+
3+
3+
4+
4+
4+
4+
Sensor distance (inches)
2
4
4
4
3.25
3.25
3.25
3.25
# of Test points
6
6
6
8
8
8
8
Avg. Inflow Vel. Range (FPM)
214-235
228-251
266-293
269-296
303-333
337-371
374-411
Correction Factor (CF)
0.95
1
0.95
1
1
1
1.08
Corrected Inflow Volume (CIV)
(Avg. velocity x CF)
203-223
228-251
253-278
269-296
303-333
337-371
404-444
HEPA Filter Leak Test Data
Air Displacement (CFM)
550
570
600
725
760
800
900
Laskin Nozzles needed
1
1
1
2
2
2
3
Theoretical aerosol conc. (ug/l)
3
25
24
23
37
36
34
30
Actual aerosol conc. (ug/l)
4
14
12
12
21
18
18
15
Supply HEPA Data
Labconco P/N
3838400
3838400
3838400
3838401
3838401
3838401
3838401
Width x Depth x Height (in.)
5
36x18x3.06
36x18x3.06
36x18x3.06
48x18x3.06
48x18x3.06
48x18x3.06
48x18x3.06
Performance (CFM)
335
335
335
445
445
445
445
Pressure Drop (in. H
2
O)
.62+/-
.1”
.62+/-
.1”
.62+/-
.1”
.5+/-
.1”
.5+/-
.1”
.5+/-
.1”
.5+/-
.1”
Exhaust HEPA Data
Labconco P/N
3838500
3838500
3838500
3838501
3838501
3838501
3838501
Width x Depth x Height (in.)
5
18x18x5.88
18x18x5.88
18x18x5.88
26x18x5.88
26x18x5.88
26x18x5.88
26x18x5.88
Performance (CFM)
266
266
266
354
354
354
354
Performance (in. H
2
O)
.25-
.35”
.25-
.35”
.25-
.35”
.2-
.37”
.2-
.37”
.2-
.37”
.2-
.37”
LED Lamp Data (2 each)
7
Labconco P/N
1297503
1297503
1297503
1297504
1297504
1297504
1297504
LED Lamps (Direct Drive)
12T8-36
12T8-36
12T8-36
15T8-48
15T8-48
15T8-48
15T8-48
Color (
o
K)
4000
4000
4000
4000
4000
4000
4000
Lumens
1200
1200
1200
1850
1850
1850
1850
Glass Type
Frosted
Frosted
Frosted
Frosted
Frosted
Frosted
Frosted
Ventus Canopy Data
Labconco Canopy P/N
3889200
3889200
3889200
3889201
3889201
3889201
3889201
Canopy Slot Area (ft
2
)
.12
.12
.12
.12
.12
.12
.12
Nominal Canopy Inflow (CFM)
50
50
50
50
50
50
50
Nominal Canopy Slot Velocity (FPM)
240-260
240-260
240-260
240-260
240-260
240-260
240-260
Canopy Vacuum (in. H
2
O)
0.15
0.15
0.15
0.15
0.15
0.15
0.15
DIM Total Exhaust Volume
C Canopy (CFM)
8
263
290
316
333
369
404
474
Traverse Total Exhaust Volume
C Canopy (CFM)
9
289
319
348
366
405
444
521
Volume of Hull
Cubic Feet
29
29
29
39
39
39
39
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...