41
Adjusting Blower Speed
During initial or annual certification of the biosafety cabinet, the blower speed may need
to be adjusted. Blower speed must only be adjusted by a trained, qualified certifier. To
adjust the blower speed, following these instructions.
1. From the Main Menu, press
[MENU]
, using
[UP]
or
[DOWN]
select the
Tools
icon. Press
[OK/MUTE]
.
2. Use
[UP]
or
[DOWN]
to select the
Password
icon, press
[OK/MUTE]
.
3. From the
Warning
Screen, press
[OK/MUTE]
to acknowledge the warning that
the following screens are restricted.
[MENU]
Tools
Password
[OK/MUTE]
WARNING !
RESTRICTED AREA
FOR TECHNICIAN
USE ONLY
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...