Diagnostics
When troubleshooting the CApture BT Fuming Chamber, it can be helpful to test
functionality of critical components individually. The Diagnostic mode allows for the
testing of individual components. To access the Diagnostic menu, on the Main Menu,
select
SETTINGS
by moving the red arrow to Settings Menu option, and press
[OK]
.
From the Settings Menu (see Fig. 6-5), select
DIAGNOSTIC
by moving the red arrow to
this sub-menu, and press
[OK]
. The Diagnostic Screen is displayed (see Fig. 7-20).
Use
[UP]
and
[DOWN]
to select a component to test, then press
[RIGHT]
to start the
test. The
NO
next to that component will change to
YES
while active. Each component
will activate for 30 seconds (when doors closed). The expected result for each
component test is listed below:
Recric. Blower
Should hear Recirculation Blower turn on. Recirculation Blower will alternate between
slower and faster speeds for 15 seconds on each speed.
Purge Blower
Should hear Exhaust Blower turn on. Exhaust Blower will run at one speed. Run this
test two (2) times in succession (no more than 5 seconds between tests) to test Exhaust
Vent Valve, which should open by end of second test.
Humidify
Should hear Recirculation Blower turn on, the Humidifier Fan will also turn on although it
is very quiet, and the Humidifier Valve will open (moderate click sound upon actuation).
Hotplate
Hot Plate heats up. CAUTION! Do
NOT
touch hot plate!
Door Locks
Both Main & CA Chamber Door lock & stay locked while active.
Figure 7-20
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...