66
Service Components
– Upper Compartment
Serviceable parts are located in the upper compartment of the CApture BT Fuming
Chamber. These components should not need to be maintained by the user, but if a
failure occurs, these components can be replaced as follows:
ALWAYS UNPLUG THE POWER CORD BEFORE SERVICING THIS PRODUCT!
1. Circuit Breaker
–
P/N 1327208 (115v: qty 1); P/N 1327209 (230v: qty 2)
Unscrew knurled nut on outside of chamber, disconnect wires.
2. LED Driver
– P/N 3181300
Remove two (2) nuts holding driver in place, disconnect all wires.
NOTE - push down on two small tabs on LED Light Strip to release wires from Light
Strip, see Fig. 7-14b above. Butt splice connectors or wire nuts (qty 4) required.
3. LED Light
– P/N 3181100
Remove two (2) nuts and screws holding the Aluminum Heat Sink Strip to the Support
Bracket, disconnect wires as shown in Fig. 7-14b above. Separate LED Light Strip from
Heat Sink Strip by removing screws & nuts.
4. Main Door Limit Switch
– P/N 3832300
Remove two (2) Nyloc Nuts holding Switch Bracket to Frame. Note wire locations on
switch for correct reassembly, then disconnect wiring. Remove two (2) screws & nuts
holding Limit Switch to Switch Bracket.
Figure 7-14
1. REMOVE T20
TORX HEAD
SCREWS (22)
2. REMOVE TOP COVER TO
EXPOSE COMPARTMENT
TOP COVER NOT SHOWN FOR
CLARITY
Figure 7-14b
4
3
PUSH
TWO
SMALL
TABS
WIRES
BACK OF UNIT
3
1
2
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...