63
Service Operations
The operations in this section provide instructions to service the CApture BT Fuming
Chamber in the event a component stops working, or a power loss occurs with evidence
in the chamber. This section also provides instructions for moving or storing the
chamber.
Manual Override of Main Door Lock
If power is lost while fuming, the Main Door will be locked. To manually override the
Lock, and remove evidence to prevent over-fuming, perform the following procedure:
CAUTION!
– CA fumes may be heavily present – Use proper PPE!
1. Open the Water Bottle Door.
2. Locate the Override Latch (see Fig. 7-12).
3. Push up on the Override Latch, and while holding the Override Latch up, open
the Main Door Handle.
If the Main Door is closed before power is restored, it will lock again. Repeat
procedure above to re-open Main Door.
Figure 7-12
MAIN
DOOR
HANDLE
OVERRIDE
LATCH
WATER
BOTTLE
DOOR