Replace Humidity Sensor (As Needed)
The RH Sensor in the CApture BT Fuming Chamber is a consumable. Eventually it will
not provide accurate data. If the RH Sensor cannot be recalibrated during annual
maintenance (see
Recalibrate Humidity Sensor (Annually)
previously in this section),
follow these steps to replace the Humidity Sensor.
1.
Open the Water Bottle Door (marked ‘H2O’ on the door label).
2. Remove the Water Bottle from its cradle and pull the Water Tube out of the
bottle. Set the Water Bottle aside.
3. Locate the Humidity Sensor behind and to the right of the Water Bottle. See Fig.
7-8.
HUMIDITY
SENSOR
HUMIDITY
SENSOR
PLATE
HARNESS
CONNECTOR
Figure 7-8