Inspect / Replace Pre-Filter (Monthly)
Visual inspect the Pre-Filter. Access the Pre-Filter as follows:
1. Locate the thumbscrew on the Pre-Filter Retainer (see Fig. 7-2). Loosen it until
the Pre-Filter Retainer hinges open. Note
– If the thumbscrew is too tight to hand
turn, a Phillips screwdriver may be used to loosen it.
If it has white residue (CA dust) built up or flaking off significantly when touched/moved,
replace the Pre-Filter as described:
1. With the Pre-Filter Retainer still open, roll the used Pre-Filter up, fold in half and
place carefully into a small trash bag.
2. Place the new Pre-Filter (orange side down) into the Pre-Filter Retainer. Verify
Pre-Filter orientation and fit within the retainer so it covers the large openings in
the Pre-Filter Retainer completely.
3. Hinge Pre-Filter Retainer closed, and re-secure the thumbscrew tightly.
Figure 7-2
THUMBSCREW
MAIN CARBON PRE-FILTER