49
Recommended Maintenance Schedule
Maintenance Frequency
Activity
Weekly
Monthly Annually
As Required
Approximate # of Cycles
25-50
50-100
250-500
500-1000
Drain Humidifier Tank
•
•
•
•
Check and Replace Pre-Filter
•
•
•
Scrape interior walls and glass to
remove CA residue
•
•
•
Inspect Recirculation Blower
Intake
•
•
•
Recalibrate Humidity Sensor
•
•
•
Replace Main Carbon Filter
•
•
Replace Humidity Sensor
•
Vacuum CA powder residue from
tubing behind Hot Plate
•
Table 7-1