41
CA Fuming
– Program Run
Once the CA Fuming Program Parameters have been established, and you are at the
ADD CA
menu screen, the fuming cycle is ready to begin. Any alerts must be resolved,
as described in
Loading Evidence
earlier in this section. And, take note of any routine
maintenance to be completed after the current cycle is finished and all evidence
removed.
It is recommended to place a test print on a small piece of material (clean plastic or
glass slides work well for this test print) inside the chamber on every Fuming Cycle to
ensure proper print development.
Before starting the fuming cycle, measure the proper amount of CA into a Tin. It is
recommended to use a calibrated balance to measure the amount of CA placed into the
tin for consistent results. Place the tin inside the CA Chamber Tin Holder (see Fig. 6-
12), close the CA Chamber Door.
NEVER PLACE CA GLUE DIRECLTY ON THE HOT PLATE!!! This can damage the
unit! If CA glue is accidentally spilled on the Hot Plate, clean it off immediately!
To begin the fuming cycle, press
[START]
.
Figure 6-12
CA CHAMBER DOOR
CA CHAMBER
TIN HOLDER
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...