MAGNETIC HOTPLATE & MAGNETIC STIRRER 

USER MANUAL 

[email protected]  

 

1800 052 226 

 

 

 

ABN 57 622 896 593 

labcoscientific.com.au

 

 

 

PO Box 5816, Brendale, QLD 4500 

 

Version 1.1 

Содержание 400.100.130

Страница 1: ...MAGNETIC HOTPLATE MAGNETIC STIRRER USER MANUAL sales labcoscientific com au 1800 052 226 ABN 57 622 896 593 labcoscientific com au PO Box 5816 Brendale QLD 4500 Version 1 1...

Страница 2: ...2 2 PROPER USE 3 3 INSPECTION 4 4 CONTROL 4 5 TRIAL RUN 6 6 USING THE EXTERNAL TEMPERATURE SENSOR 6 7 RESIDUAL HEAT WARNING HOT 7 8 LOGGING FUNCTION 7 9 FAULTS 7 10 MAINTENANCE AND CLEANING 8 11 ASSOC...

Страница 3: ...e that only trained staff works with the instrument Risk of burn hotplate model Caution when touching the housing parts and the hotplate they can reach temperature of 500 C Pay attention to the residu...

Страница 4: ...the user manual Accessories must be securely attached to the device Always disconnect the plug before fitting accessories When using the external temperature sensor the tip of the measuring sensor mus...

Страница 5: ...NOB The heating function is switched SPEED CONTROL KNOB Set the rated rotary speed The stirring function is switched ON or OFF by LED HEAT When the heating function is switched ON the Led Heat is lit...

Страница 6: ...rror is occurring SD Displays when using SD cards RATED VALUE REAL VALUE Displays value in case of heating or stirring function switching ON Note If both heating and stirring functions have been start...

Страница 7: ...nsfer the vessel when the instrument is powered on 6 USING THE EXTERNAL TEMPERATURE SENSOR The external temperature sensor PT1000 is a standard accessory If the sensor is plugged in Probe will be disp...

Страница 8: ...n the left side of the panel After inserting it turn on the power switch and operate the heating as described above Set temperature at 1 minute intervals The current temperature will be logged automat...

Страница 9: ...uipment during cleaning of the instrument Send in the case of service the instrument back in the packaging carton Storage packing is not sufficient for despatch Please use additional suitable packagin...

Страница 10: ...TE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection again...

Страница 11: ...0 ROTARY SPEED DISPLAY LCD PLATE MATERIAL Stainless steel DIMENSIONS OF WORKPLATE W X D 254 x 254mm HEATING POWER W 500 TEMPERATURE RANGE C RT 340 increment 1 TEMPERATURE DISPLAY C LCD TEMPERATURE DIS...

Страница 12: ...RIPTIONS 400 100 130 Stirrer Magnetic Hotplate up to 30L ACCESSORIES 400 100 215 Sensor Temperature PT1000 100mm 400 100 218 Sensor Temperature PT1000 230mm 400 100 219 Sensor Temperature PT1000 230mm...