Instruction Manual for Model 8112 BNC

INTRODUCTION 

Model 8112BNC is a small sized, high sensitivity clamp – on AC current probe 
designed for use with an oscilloscope. The differential coil incorporated permits 
a wide range of current measurements with the minimum effect of the external 
magnetic field. Jaws can be opened or closed simply by sliding the knob. 
Three current ranges can be switch selected. 

SPECIFICATIONS 

Range Switch 

Position 

Input 

Output   

Accuracy 

Frequency 

Range 

200mA/200mV

(1mA/mV) 

0-0.5A AC 

0-500mV AC 

2% of output 
0.2mV 

100Hz-1kHz 

0-1A AC 

0-1000mV AC 

4% of output 
0.4mV 

50Hz-10kHz 

2A/200mV 

(10mA/mV) 

0-20A AC 

0-2000mV AC 

1% of output 
0.2mV 

40Hz-1kHz 

1.5% of 

output 0.2mV

1kHz-10kHz 

20A/200mV 

(100mA/mV) 

0-20A AC 

0-200mV AC 

1% of output 
0.1mV 

40Hz-1kHz 

1.5% of 

output 0.2mV

1kHz-10kHz 

20-40A AC 

200mV-400mV 

AC 

2.5% of 

output 

50Hz-10kHz 

40-80A AC 

400mV-800mV 

AC 

2.5% of 

output 

100Hz-10kHz 

Output Impedance – 2k ohms 

Frequency Characteristics – Within 

3dB at 30Hz to 100kHz 

Size of conductor under test
    8mm maximum 

Output cord
    Approx. 1.5mm long coaxial cable with a connector 

Accessory
    Pouch 

MEASUREMENT METHOD 

         

(1) Jaws          (2) Knob      (3) Range Selector                                      (4) Connector 

                                                                            Fig.1 
(1) Connect the connector (4) of the adaptor to the vertical input terminal 

of the oscilloscope. 

(2) Slide the knob (2) backward to open the jaws and clamp on a current 

carrying conductor (refer to Fig.2). In example (a), output voltage 
selected by the range selector will be obtained. In case of (b), output voltage   
twice the value selected by the range selector will be obtained since two 
turns of the conductor is clamped as illustrated. 

(3) Arrow mark on the jaws indicates polarity relative to the current flowing 

through the conductor. When the current flows in the arrowed direction, output 
will be positive (+). 

(4) Select a desired range according to the current flowing through the conductor. 
(5) Where the current waveform is affected by noise, clip the earth lead onto 

the earth terminal of the probe and connect it to the circuit under test. 