Delay
PT5040 Operation Manual Chapter 6 Introduction to [SYSTEM]
MEAS STOP soft key is pressed.
BUS
TRIGGER commands are sent to PT50xx via RS232SC, USBTMC,
USBCDC or LAN interface.
Time
Delay time refers to the time between the end of a measurement and the start of the next
measurement. The delay function is only available when trigger mode is set to INT trigger mode.
The trigger delay time can be set from 0.1s to 99s by 0.1 steps.
Move the cursor to this zone. User can input the delay time by pressing numerical keys and
ENTER for confirmation. The input unit is s. Following soft keys can also be used to set the
delay time.
They are coarse adjustment keys used to change the delay time by 1.0s.
They are fine adjustment keys used to change the delay time by 0.1s.
BDV Test
BDV test is a test method for turn-to-turn insulation test. Starting from Start Volt and ending at
End Volt by a resolution of Volt Step, the test is finally done by comparing with four comparison
methods. Move the cursor to this zone, following soft keys will be displayed.
ON Turn BDV test on.
OFF Turn BDV test off.
Start Volt
Start Volt ranges from 100V to 3000V(or 5000V) but cannot be larger than End Volt. Move the
cursor to this zone, following soft keys will be displayed.
They are coarse adjustment keys used to change Start Volt by 100V.
They are fine adjustment keys used change Start Volt by 10V.
End Volt
End Volt ranges from 100V to 3000V(or 5000V) but must be larger than Start Volt. Move the
cursor to this zone, following soft keys will be displayed.
They are coarse adjustment keys used to change End Volt by 100V.
They are fine adjustment keys used change End Volt by 10V.
5-6