6.4.7
Temperature unit
Here you can choose the desired temperature units Degree Celcius (°C)
or Fahrenheit (°F). Factory setting: “°C”
6.4.8
Language selection
Here you can choose the desired language for the menus of the
i‑CON VARIO 2. Factory setting: “English”
6.4.9
Adjustable shut‑down function (idle state)
The adjustable shut‑down function (Idle state) serves the purpose of
energy saving and prolonging the lifetime of soldering tip and heating
element. In idle state the tools cool down to room temperature. The
bottom line in the display then shows the text [Idle state]. You can end
idle state at any time by actuating the i‑OP.
Please proceed as follows to activate the function [Idle state]:
■
Open the configuration mode and choose the option [Idle state].
■
Activate the menu [Idle state] by turning the i‑OP.
■
Press the i‑OP to activate the cursor, pressing the i‑OP again de‑
activates the cursor.
■
Enter the desired time from [010] to [240] minutes by turning the
i‑OP accordingly. The input [000] deactivates the function [idle state].
With the cursor activated, turn the i‑OP to change the value of the parameter
accordingly. The set time applies for both connected tools. The set time starts to run,
only after both tools are in [Standby]. If the tools are still not used during this time, idle
state is activated after this time has expired.
With the i‑TOOL and the CHIP TOOL VARIO idle state is automatically terminated when
the tool is taken out of the holder. Once the first tool has “woken up” from idle state,
the second tool will also be reactivated. The i‑TOOL AIR S must be explicitly switched
on. Find more about this in chapter 6.2.4 [Working with the i‑TOOL AIR S].
A controlled heating plate is not automatically activated! Find out more about this in