Technical Data
Measurement range
→
Accuracy
1)
XCG-xx-x
A
-1
XCG-xx-x
B
-1
≤1% of reading
≤0.5% of reading
Temperature effect on zero
0.02% F.S./ °C
Temperature effect on span
0.02% of reading/ °C
Resolution
0.05% F.S.
Gas type dependence
none
Output signal analog
(measuring signal)
Voltage range
–0.2 … +10.24 V
Measuring range
0 … +10 V
Relationship voltage-pressure
linear
Output impedance
10
Ω
(short-circuit proof)
Loaded impedance
>10 k
Ω
Response time
100 ms
Gauge identification
Resistance 13.2 k
Ω
refer-
enced to supply common
1)
Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C
ambient operating temperature without temperature effects after
operation of 2 h.