Quick Start OPTIFLUX 6000

 

OPTIFLUX 6000 

 

Limits for operating pressure and product temperature 

 

Size of connection 

Max. process 

temperature 

Type of connection 

mm inch 

Max. operating pressure 

in bar (psig) 

°C / °F 

DN 10* - 40 

40 (580) 

140 / 284 

DN 50 - 80 

25 (360) 

140 / 284 

DN 100 

16 ( 230) 

140 / 284 

Aseptic weld on for pipes to DIN 
11850  ** 

DN 125 – 150 

10 (145) 

140 / 284 

12* - 38 

40 (580) 

140 / 284 

51 – 76,1 

25 (360) 

140 / 284 

101,6 

16 ( 230) 

140 / 284 

Aseptic weld on for pipes to ISO 
2037 

114,3 – 139,3 

10 (145) 

140 / 284 

DN 10* - 40 

40 (580) 

140 / 284 

DN 50 – 80 

25 (360) 

140 / 284 

DN 100 

16 ( 230) 

140 / 284 

Dairy screw to DIN 11851 

DN 125 – 150 

10 (145) 

140 / 284 

Screwed to SMS 1145 

25 – 100 

6 (90) 

140 / 284 

DN 25 – 40 

40 (580) 

140 / 284 

DN 50 – 80 

25 (360) 

140 / 284 

DN 100 

16 ( 230) 

140 / 284 

Flanges to DIN 11864-2A 

DN 125 – 150 

10 (145) 

140 / 284 

12* - 51 

16 ( 230) 

120 / 248 

63.5 – 76,1 

10 (145) 

120 / 248 

100 - 

(115)  120 / 248 

Clamp joint to ISO 2852 

125 - 150 

5 (72) 

120 / 248 

DN 25 - 50 

16 (230) 

140 / 284 

Clamp joint to DIN 32676 ** 

DN 65 - 100 

10 (145) 

140 / 284 

½ - 3 

20.5 (295) 

120 / 248 

Clamp joint to TRI-CLOVER 

13.8 (200) 

120 / 248 

Vacuum load 

all versions and sizes 

0 mbar abs. (0 psia) 

 

*   for meter size DN 2,5…DN 10 

 

 

  

** separate flow sensor max. temperature up to 150 

º

C / 302 

º

F  

 

 

Pressure rating and max. torque for 11864-2A flange version 

 

 

 

Nominal size         
DN mm 

Pressure 
rating 

Bolts max. 

Torque      

Nm 

25 

PN 40 

4 × M 6 

40 

PN 40 

4 × M 8 

16 

50 

PN 25 

4 × M 8 

16 

80 

PN 25 

6 × M 8 

16 

100 

PN 25 

6 × M 8 

16 

125 

PN 10 

6 × M 10 

32 

150 

PN 10 

6 × M 10 

32 

 
1)

 

The max. allowable torque is dependent on the 

gasket material, see technical data sheet for detailed 

information. 

2)

 

Bolts to be used from material SS, class 70-A2. 

 

 

Limits / Vacuum load 

For general limits see instrument type plate! 
Vacuum load only for PFA-lined flowmeters; see manual for detailed 

information. 

 

Max. Torque 
Step 1: approx. 50% of max. torque 

Step 2: approx. 80% of max. torque 

Step 3: 100% of max. torque given in tables above 

Содержание OPTIFLUX 6000

Страница 1: ... supplied in a separate document that describes all hazardous area relevant information Responsibility as to suitability and intended use of this device rests solely with the user The supplier does not accept any liability resulting from misuse by the customer Improper installation and operation may lead to loss of warranty In addition the General conditions of sale found on the back of the invoic...

Страница 2: ...Quick Start OPTIFLUX 6000 2 OPTIFLUX 6000 Handling and installation instructions ...

Страница 3: ...C 300 40 C 140 C IFC 300 40 F 284 F PFA 40 C 180 C IFC 010 25 C 120 C PFA 40 F 356 F IFC 010 13 F 248 F Ambient temperatures Ambient temperatures C min max F min max IFC 300 40 C 65 C IFC 300 40 F 149 F IFC 010 25 C 60 C IFC 010 13 F 140 F Note The min max temperatures for the compact devices are dependent on the liner converter combination and the used type of adapter See data sheet ...

Страница 4: ...Quick Start OPTIFLUX 6000 4 OPTIFLUX 6000 Special considerations ...

Страница 5: ...m the adapters by loosen the screws as well as the L shaped gaskets 3 Weld the adapters completely to the pipe 4 After the pipe has cooled down reinstall the gasket and mount the flowmeter Grounding Metal pipelines not internally coated Grounding without grounding rings Metal pipelines with internal coating and internally non conductive pipelines Grounding with grounding rings Grounding rings Clea...

Страница 6: ... 8 115 120 248 Clamp joint to ISO 2852 125 150 5 72 120 248 DN 25 50 16 230 140 284 Clamp joint to DIN 32676 DN 65 100 10 145 140 284 3 20 5 295 120 248 Clamp joint to TRI CLOVER 4 13 8 200 120 248 Vacuum load all versions and sizes 0 mbar abs 0 psia for meter size DN 2 5 DN 10 separate flow sensor max temperature up to 150 ºC 302 ºF Pressure rating and max torque for 11864 2A flange version Nomin...

Страница 7: ...7 Quick Start OPTIFLUX 6000 OPTIFLUX 6000 ...

Страница 8: ...al Area H Block Pimpri Poona 411018 TEL 91 0 202 7442020 FAX 91 0 202 7442020 e mail pcu vsnl net Iran KROHNE Liaison Office North Sohrevardi Ave 26 Sarmad St Apt 9 Tehran 15539 TEL 98 21 874 5973 FAX 98 21 850 1268 e mail krohne krohneiran com Italy KROHNE Italia Srl Via V Monti 75 I 20145 Milano TEL 39 0 2 4 30 06 61 FAX 39 0 2 43 00 66 66 e mail info krohne it Korea KROHNE Korea Room 508 Miwon ...

Отзывы: