1.3.4  Verify successful panel configuration 

1.  Launch Ultra. 

2.  Select a test patient. 

3.  Click 

Acquire

 to open the acquisition 

screen and verify panel status in the upper 
left corner. 
The status transitions from “Changing 
Power Modes” to “Connected” (Figure 17). 
 

 

Figure 17 

The status, “Connected”, indicates that the 
panel is configured correctly. 

 

IMPORTANT

: Keep the Panel CD in a safe place. 

The CD is needed to install DR panel if there 
should be PC hardware or software catastrophic 
failure. 

 
 

Note: To Power off the panel. Flip the switch on the 
Power Box (XG-IFBox) to the off position. The panel will 
power off after 60 seconds. Switching back on at any 
time after the panel is turned off only requires the 30 
second start up and initialization, as instructed in Figure 
5. 

 

1.4  KDR Panel Calibration 

1.4.1  Before you begin 

1.  Remove the grid. 

2.  Adjust the SID to 72 inches. 

3.  Ensure that the collimator light is at least 1 

inch past the edges of the detector for full 
coverage. 

4.  Verify that nothing is in the X-Ray field. 

1.4.2  KDR panel calibration 

1.  To begin calibration, exit Ultra and return to 

the desktop. 

2.  Browse to C:\Program Files (x86)\Sedecal 

TechService and double click on 

TechService.exe

,  

3.  Select 

Manual Calibration

4.  Select 

WS1

 to configure workstation 1 

(Bucky 1). 

5.  Select 

Go to User Mode

6.  Navigate to 

C:\opal\plugins32\Aero

7.  Double-click the file, 

AerosdkGainCalibrationSeq_Triggered.exe

A command line window, Aero SDK Library 
Sample Program, is displayed. 

Содержание Ultra KDR

Страница 1: ...Ultra KDR Configuration and Calibration Guide...

Страница 2: ...el Registration Configuration and Calibration 3 1 1 Before you begin 3 1 2 Configuring the KDR panel 4 1 3 Panel Registration 4 1 3 1 Before you begin 4 1 3 2 Panel registration 5 1 3 3 Panel configur...

Страница 3: ...n Manual A9KXBB01EN01_160316 Before starting panel setup verify the network settings for the onboard card The following configuration applies to Ultra 4 1 0 x or later with the KDR Panel and driver 10...

Страница 4: ...cessing Default value 0 X Ray Synchronization Integrated XG IF Box ID 1 Exposure Time 1 0 Figure 3 Device configuration 1 3 Panel Registration 1 3 1 Before you begin You will need to power on the KDR...

Страница 5: ...3 On the Device Configuration tab under Device Identity section enter the Panel serial number Figure 6 You can find this number located on the back of the panel Figure 6 Device configuration 4 Click...

Страница 6: ...n the Device Identity list form in the IF cable packaging Figure 9 item 2 8 Click OK Figure 9 item 3 The KONICA MINOLTA AeroSDK Setting Tool main screen will become active 9 On the KONICA MINOLTA Aero...

Страница 7: ...ng Figure 13 Figure 13 Konica Minolta KDR CD 1 3 3 Panel configuration 1 Insert the installation CD The Aero Installer window is displayed Figure 14 Select the Browse To option Highlighted in yellow a...

Страница 8: ...to the off position The panel will power off after 60 seconds Switching back on at any time after the panel is turned off only requires the 30 second start up and initialization as instructed in Figu...

Страница 9: ...h Service adjust techniques to 80 KvP 10ms and 100 mA 14 When ready take your first exposure 15 If the Recommended Technique is low or high The application will prompts you to adjust your technique by...

Отзывы: