396
10 .2 Error codes and countermeasures
10 .2 .4
AeroDR Interface Unit2/AeroDR Generator Interface Unit/
AeroDR Generator Interface Unit2/AeroDR Portable RF Unit/
AeroDR Portable RF Unit2 errors
Error codes starting with "G" are indicating AeroDR Interface Unit2, AeroDR Generator Interface Unit, AeroDR
Generator Interface Unit2, AeroDR Portable RF Unit, AeroDR Portable RF Unit2 errors .
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• Error codes are common to the AeroDR Interface Unit2, AeroDR Generator Interface Unit, AeroDR Generator Interface
Unit2, AeroDR Portable RF Unit, and AeroDR Portable RF Unit2 .
•
[Device name] under the Displayed message column and the "corresponding device" under the Problems/countermeasures
column indicate a device in which an error is occurring .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Error code
Displayed message
Problems/countermeasures
G0600X
Version up failed .
Contact a service representative .
Contact Konica Minolta technical representatives .
G0601X
Restart the AeroDR Interface Unit / AeroDR Interface
Unit2, and CS-7 .
If the error cannot be released, contact a service rep-
resentative .
(1) Restart the corresponding device and this device .
(2) Contact Konica Minolta technical representatives .
G0602X
Restart the AeroDR Interface Unit / AeroDR Interface
Unit2, and CS-7 .
If the error cannot be released, contact a service rep-
resentative .
G06030
Restart the AeroDR Interface Unit / AeroDR Interface
Unit2, and CS-7 .
If the error cannot be released, contact a service rep-
resentative .
G2010S
G2011S
[Device name] There is a startup error.
Release the exposure switch.
Initial status error of the corresponding device oc-
curred .
Release the exposure switch.
G2012S
[Device name] There is a startup error.
Contact a service representative .
Contact Konica Minolta technical representatives .
G2020S
G2021S
[Device name] There is a startup error.
Release the exposure switch.
Initial status error of the corresponding device oc-
curred .
Release the exposure switch.
G2022S
[Device name] There is a startup error.
Contact a service representative .
Contact Konica Minolta technical representatives .
G2030S
G2031S
[Device name] There is a startup error.
Release the exposure switch.
Initial status error of the corresponding device oc-
curred .
Release the exposure switch.
G2032S
[Device name] There is a startup error.
Contact a service representative .
Contact Konica Minolta technical representatives .
G22000
[Device name] There is a X-ray device exposure link
error . Restart the X-ray device and AeroDR Interface
Unit/AeroDR Interface Unit2 .
Link error between the X-Ray device and the corre-
sponding device occurred .
Restart the X-ray device and the corresponding de-
vice .
G2201U
[Device name] There is a X-ray device exposure link
error . Contact a service representative .
Contact Konica Minolta technical representatives .
G2300U -
G2302U
[Device name] There is an internal device error or a
communication error . Restart the AeroDR Detector,
AeroDR Interface Unit/AeroDR Interface Unit2 and
CS-7 .
System error of the corresponding device occurred .
Restart the corresponding device, and the device .
G2303U
[Device name] Exposure link failed, X-rays are not
exposed. Please take exposure once again.
X-rays were not irradiated. Try performing the expo
-
sure again .
The release of the exposure switch may occur quickly,
so exposure while making sure to release the expo
-
sure switch properly .
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...