9.1 Various problems and countermeasures
377
Chapter 9
6
The confirmation screen is displayed on
this unit connected to the same system .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• After the switchover of JM, contact Konica Minolta
technical representatives as soon as possible .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• Switching to backup JM may take about 1 minute .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
9 .1 .2
Reacquisition of images
from the DR Detector
Reacquire the images when images cannot be re-
ceived from the DR Detector after the exposure due to
reasons such as communication errors, etc .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• Reacquire images that have not been transferred us-
ing this device, in which a communication error has
occurred .
•
If you end your examination while an image that has
not been transferred is left in the DR Detector, you will
not be able to acquire the image again .
• When images that have not been transferred remain in
the DR Detector, exposure cannot be performed with
that DR Detector .
• If an image cannot be received using the Auto stitch-
ing exposure with DR Detector, cancel the image
reacquisition and do not perform any changes to the
exposure device.
• In the event of a communication error and depending
on the timing at which the error has occurred, images
may not be displayed and the DR Detector may not
contain any images that have not been transferred
(the same status as that observed when no exposures
have been carried out) . In such a case, a message
may be displayed . Suspend the study and display the
exposure screen from the list screen. This operation
may successfully recover the images .
• Non-transferred images need to be reacquired be-
cause the One Shot stitching exposure conducts the
combining and displaying images after acquiring all
the images .
• Delete the non-transferred images from all the DR
panels used in the exposure, to delete the non-
transferred images by the reacquiring of the One Shot
stitching exposure.
(a) Press [OK] when a message "Unreceived image
of One Shot Stitching exposure is deleted. Unre
-
ceived image of same exposure is present in other
panels ." appeared .
(b) Press the framed device icon .
(c) Press [OK] when a confirmation screen appeared.
(d) Press [OK] when the confirmation screen appeared
again .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...