E-52
N
ote
To make pass/fail judgment during measurement, it is necessary to set tolerances for the target color to
be used. For details, refer to “Setting Color Difference Tolerances” (page E-52).
Deleting a Color Difference Target Data
There are two methods for deleting target color data. One is to delete the target color data set in “T
*
” of
the measurement conditions (COND 1 to COND 6) one by one. The other is to delete all the target color
data set in the instrument.
N
ote
• When target color data is deleted, its tolerances will also be deleted.
• You will not be able to view the color difference for measurement data that used the deleted target color data.
[Setting Procedure]
Display the target color data to be deleted, as explained in “Setting a Color Difference Target
Data”.
1
Turn
B
to select “DEL”, then press
A
.
2
Turn
B
to select the desired deletion method, then press
A
.
Deletion Method
• T
*
DELETE:
Deletes the currently displayed target color data.
• If the target color data to be deleted by another “COND
*
” or “TASK
*
” has been selected, a message
“THIS TARGET IS USED IN OTHER COND OR TASK. DO YOU DELETE IT?” will appear. Se-
lect “YES” if you want to delete or select “NO” if not, using
B
, then press
A
.
• The target color data nos. next to that of the deleted target color data will scroll up automatically.
• ALL DELETE:
Deletes all the target color data currently set in the instrument. All the target color data will be deleted
without a confirmation message.
N
ote
If the target mode is “linked to each data.”, “T---” is displayed for target color number if the target color linked
at the time of measurement is deleted, and the color difference for that data cannot be displayed.
Содержание CM-2600d
Страница 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Страница 11: ...E 9 Chapter 1 Before Using the Instrument...
Страница 27: ...E 25 Chapter 2 Preparation for Measurement...
Страница 62: ...E 60 Setting Color Difference Tolerances...
Страница 63: ...E 61 Chapter 3 Measurement...
Страница 79: ...E 77 Chapter 4 Other Functions...
Страница 108: ...E 106 TASK Mode...
Страница 109: ...E 107 Chapter 5 Troubleshooting...
Страница 114: ...E 112 Troubleshooting...
Страница 115: ...E 113 Chapter 6 Appendix...
Страница 122: ...E 120 Dimensions Dimensions 193 96 96 69 mm The measurement area selector is not available with the CM 2500d...
Страница 123: ...E 121 Dimensions...
Страница 126: ......
Страница 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...