Observer/Illuminant 1
Select an observer angle of either 2° or 10° and the illuminant used to measure colorimetric value.
Memo
Observer/Illuminant 1 is set to “10°/D65” when the instrument is shipped from the factory.
Operating Procedure
Start the procedure from the <Measurement cond.> - <Observer/Illuminant> screen.
1
Use [
▲
] or [
▼
] to move the cursor to “Observer/
Illuminant 1”, and then press the [Confirmation] key.
The <Observer/Illuminant 1> screen
is displayed.
The scroll bar on the right of the screen shows that there
are additional observer/illuminant settings that cannot fit
on the screen.
2
Use [
▲
] or [
▼
] to move the cursor to the desired item.
Moving the cursor past the top or bottom of the list will show
additional observers/illuminants that cannot fit on the screen.
Settings
{
2°: 2° observer angle (CIE 1931)
{
10°: 10° observer angle (CIE 1964)
and
{
A
: Standard illuminant A (Incandescent lamp,
Color temperature: 2856K)
{
C
: Illuminant C (Daylight, Small relative value of
spectral distribution in the ultraviolet region;
Color temperature: 6774K)
{
D65 : Standard illuminant D
65
(Daylight, Color
temperature: 6504K)
{
D50 : Standard illuminant D
50
(Daylight, Color
temperature: 5003K)
{
ID65 : Indoor daylight illuminant ID65 (Daylight after passing
through window glass, Color temperature: 6504K)
{
ID50 : Indoor daylight illuminant ID50 (Daylight after passing
through window glass, Color temperature: 5003K)
{
F2
: Cool white (fluorescent lamp)
{
F6
: Cool white (fluorescent lamp)
{
F7
: Color rendering A daylight natural (fluorescent lamp)
{
F8
: Color rendering AAA natural white (fluorescent lamp)
{
F10 : 3-band type natural white (fluorescent lamp)
{
F11 : 3-band type cool white (fluorescent lamp)
{
F12 : 3-band type warm white (fluorescent lamp)
Содержание CM-23d
Страница 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Страница 30: ...27...
Страница 33: ...30 Measurement 2 Attach the Wrist Strap Attaching the Wrist Strap Wrist strap attaching position Preparation...
Страница 82: ...79...
Страница 111: ...108 Setting 3 MEMO...
Страница 112: ...109...
Страница 132: ...129...
Страница 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Страница 138: ...135...
Страница 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Страница 142: ...139 Dimensions Unit mm...
Страница 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...