30
Specifications of Each Probe
CA-P427
CA-P427H
Measurement area
Ø27 mm
Ø27 mm
Acceptance angle
±2.5°
±2.5°
Accuracy guaranteed measurement distance
30±10 mm
30±10 mm
Display range Luminance
0.0001 to 5000 cd/m
2
0.0001 to 30000 cd/m
2
Chromaticity
Displayed in 4 digits
Displayed in 4 digits
Luminance
Accuracy guaranteed luminance range
0.001 to 5000 cd/m
2
0.01 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.001 - cd/m
2
±9%
---
0.01 - cd/m
2
±2%
±9%
0.1 - cd/m
2
±1.5%
±2%
1 - cd/m
2
±1.5%
±1.5%
10 - cd/m
2
±1.5%
±1.5%
100 - cd/m
2
±1.5%
±1.5%
Repeatability
(2
σ
)
*1
AUTO
0.001 - cd/m
2
10%
---
0.01 - cd/m
2
1%
10%
0.1 - cd/m
2
0.40%
1%
1 - cd/m
2
0.10%
0.40%
10 - cd/m
2
0.10%
0.10%
100 - cd/m
2
0.10%
0.10%
Chromaticity Accuracy guaranteed luminance range
0.01 to 5000 cd/m
2
0.1 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.01 - cd/m
2
±0.003
---
0.1 - cd/m
2
±0.002
±0.003
1 - cd/m
2
±0.002
±0.002
10 - cd/m
2
±0.002
±0.002
100 - cd/m
2
±0.002
±0.002
In 100 cd/m
2
(for monochrome)
*2
100 cd/m
2
±0.003
±0.003
Repeatability
(2
σ
)
*1
AUTO
0.01 - cd/m
2
0.0035
---
0.1 - cd/m
2
0.0015
0.0035
1 - cd/m
2
0.0004
0.0015
10 - cd/m
2
0.0003
0.0004
100 - cd/m
2
0.0002
0.0003
Flick
er (CA
-310 Mode)
*6
Flicker
(Contrast)
Measurement luminance range
5 to 1500 cd/m
2
30 to 9000 cd/m
2
Range to be measured (flicker frequency)
0.25 to 65 Hz
0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave ± 0.4%
± 0.4%
60 Hz, AC/DC 10% sine wave ± 0.7%
± 0.7%
Repeatability (2
σ
)
20 to 65 Hz,
AC/DC 10% sine wave
0.3%
0.3%
Flicker
(JEITA)
Measurement luminance range
5 to 1500 cd/m
2
30 to 9000 cd/m
2
Range to be measured (flicker frequency)
0.42 to 65 Hz
0.42 to 65 Hz
Accuracy
30 Hz, AC/DC 4% sine wave ±0.35 dB
±0.35 dB
30 Hz, AC/DC 1.2% sine wave ±0.35 dB
---
Repeatability (2
σ
)
30 Hz, AC/DC 4% sine wave 0.1 dB
0.1 dB
30 Hz, AC/DC 1.2% sine wave 0.3 dB
---
Specifications of Each Probe (1)
Содержание CA-410 Series
Страница 12: ......
Страница 40: ......
Страница 139: ......
Страница 140: ...En 9222 AA1J 12 2017 KONICA MINOLTA INC BIFBDA...